Skip to content

NA 288202Recalls & reported problems

Medos International Sàrl · Latarjet EXPERIENCE, Coracoid Top Hat Drill · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10886705026807), received between Feb 14, 2020 and Jun 3, 2026 — classified as 1 injury, 14 malfunctions.

The most frequently reported specific device problem is “Degraded” (9 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2017Class IITerminated

    Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw

    Z-2617-2017 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Degraded9
  • Crack2
  • Device Contaminated During Manufacture or Shipping1
  • Material Deformation1
  • Material Twisted/Bent1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • No Consequences Or Impact To Patient1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2026Crack

    IT WAS REPORTED THAT DRILL BIT ON THE CORACOID TOP HAT DRILL HAD A CRACK IN THE TIP. THE EVENT OCCURRED DURING HANDLING AND PRIOR TO INSERTION. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25371894 ↗
  • MalfunctionJun 3, 2026Crack

    IT WAS REPORTED THAT DRILL BIT ON THE CORACOID TOP HAT DRILL HAD A CRACK IN THE TIP. THE EVENT OCCURRED DURING HANDLING AND PRIOR TO INSERTION. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25371893 ↗
  • MalfunctionDec 5, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE UNSPECIFIED SURGICAL PROCEDURE THE CORACOID TOP HAT DRILL DEVICE HAD BONE IN IT. IT IS UNKNOWN WHETHER ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23727370 ↗
  • MalfunctionAug 25, 2025Degraded; Degraded

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT WHILE USING THE CORACOID TOP HAT DRILL DEVICE, SHARP CURVED OSTEOTOME DEVICE, LATARJET GLENOID LONG DRILL BIT 3.2 DEVICE AND THE COMBO SCREWDRIVER DEVICE IT WAS OBSERVED THAT THE DEVICES WERE BLUNT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 3 OF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22891152 ↗
  • MalfunctionNov 13, 2024Degraded; Degraded

    IT WAS REPORTED FROM AUSTRALIA THAT TWO SETS OF DEVICES HAD BLUNT INSTRUMENTS. IT WAS REPORTED THAT THE TWO CORACOID TOP HAT DRILL DEVICES, AND TWO SHARP CURVED OSTEOTOME DEVICES WERE BLUNT. IT WAS NOT REPORTED IF THE EVENT OCCURRED DURING A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 3 OF 4 FOR THE SAME EVENT IN (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20677337 ↗
  • MalfunctionNov 13, 2024Degraded; Degraded

    IT WAS REPORTED FROM AUSTRALIA THAT TWO SETS OF DEVICES HAD BLUNT INSTRUMENTS. IT WAS REPORTED THAT THE TWO CORACOID TOP HAT DRILL DEVICES, AND TWO SHARP CURVED OSTEOTOME DEVICES WERE BLUNT. IT WAS NOT REPORTED IF THE EVENT OCCURRED DURING A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 4 FOR THE SAME EVENT IN (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20677277 ↗
  • MalfunctionSep 26, 2024Degraded; Degraded

    IT WAS REPORTED BY THE SALES REP FROM CANADA THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON AN UNKNOWN DATE THE CORACOID TOP HAT DRILL DEVICE TOP HAT DRILL WAS DULL. IT WAS UNKNOWN IF A SPARE DEVICE WAS USED TO COMPLETE THE PROCEDURE OR IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. IT WAS UNKNOWN IF A THERE WAS A DELAY TO THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20314495 ↗
  • MalfunctionAug 26, 2024Degraded; Degraded

    IT WAS REPORTED FROM THE SALES REP THAT DURING DEVICE INSPECTION IT WAS OBSERVED THAT THE CORACOID TOP HAT DRILL DEVICE WAS BLUNT PRIOR TO USE. THIS EVENT DID NOT OCCUR DURING SURGERY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 2 OF THE SAME EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20071374 ↗
  • MalfunctionDec 6, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE SALES REP IN AUSTRALIA THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2023 THE CORACOID TOP HAT DRILL DEVICE SEEMED BENT. IT WAS REPOSITIONED IN THE CHUCK BUT IT DIDN'T HELP. THE DRILL IS DEFINITALY BENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18276858 ↗
  • MalfunctionApr 19, 2023Degraded; Degraded

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE CORACOID TOP HAT DRILL DEVICE WAS BLUNT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16772755 ↗
  • MalfunctionMar 8, 2023Degraded; Degraded

    THIS IS REPORT 1 OF 3 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING AN UNKNOWN ARTHROSCOPIC PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE CORACOID TOP HAT DRILL DEVICE WAS BLUNT. IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16504399 ↗
  • MalfunctionOct 7, 2022Degraded; Degraded

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN AUSTRALIA THAT DURING A POSTERIOR BONE BLOCK PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE CORACOID TOP HAT DRILL DEVICE WAS BLUNT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15564535 ↗
  • MalfunctionMar 28, 2022Mechanical Jam; Mechanical Jam

    IT WAS REPORTED BY THE AFFILIATE IN AUSTRALIA THAT DURING A LACERATION CLOSURE PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE K-WIRE GOT STUCK IN CANNULATION ON THE CORACOID TOP HAT DRILL DEVICE. THE K-WIRE WAS REMOVED WITH COCKER TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13931684 ↗
  • InjuryMay 7, 2021Degraded; Degraded

    IT WAS REPORTED BY THE SALES REP IN AUSTRALIA THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE CORACOID TOP HAT DRILL DEVICE WAS BLUNT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 11791804 ↗
  • MalfunctionFeb 14, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING AN UNKNOWN PROCEDURE THE CORACOID TOP HAT DRILL WAS SLIGHTLY BENT, AND WHEN PERFORMING THE PROCEDURE IT WAS ECCENTRIC DRILLING OF THE HOLE. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9712658 ↗

FDA clearance and classification

The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 288202

Has the NA 288202 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-2617-2017 (Class II), began on May 12, 2017: Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the NA 288202?

The FDA's MAUDE database lists 15 adverse event reports for this device, received between Feb 14, 2020 and Jun 3, 2026. In those reports, the most common specific device problems are “Degraded” (9), “Crack” (2) and “Device Contaminated During Manufacture or Shipping” (1); the most common patient problem is “Unspecified Tissue Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 288202?

None of the reports are classified as deaths, 1 as an injury and 14 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 288202 FDA approved?

Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Medos International Sàrl.

Who makes the NA 288202, and is it still sold?

The FDA's device database lists Medos International Sàrl as the company responsible for it (brand NA, model 288202). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates