Skip to content

INTRAFIX 254822Recalls & reported problems

Medos International Sàrl · INTRAFIX Advance Tibial Fastener System Screw 9 x 30mm PEEK · Class 2 · Implant

All models of INTRAFIX (32)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886705027484), received between Jun 15, 2018 and Jan 29, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration or Expulsion of Device2
  • Break1
  • Delivered as Unsterile Product1
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Muscular Rigidity1
  • No Consequences Or Impact To Patient1
  • Tissue Damage1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2026Delivered as Unsterile Product; Packaging Problem

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT PROCEDURE, THE OUTER PACKAGE OF THE INTRAFIX ADV PEEK SCW 9X30MM DEVICE WAS OPENED AND INNER SEAL APPEARED TO BE PUNCTURED. THE FILM AROUND THE SCREW HAS A CLEAR TINY HOLE IN IT WHICH REDUCED OUR CONFIDENCE IN ITS STERILITY. FROM WHAT I COULD TELL THERE WAS NO VISIBLE DAMAGE TO THE BOX. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24215459 ↗
  • MalfunctionDec 10, 2021Break; Break

    IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE TIP ON THE INTRAFIX ADV PEEK SCW (B)(4) DEVICE BROKE AS IT POSSIBLY WENT OFF AXIS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12975791 ↗
  • InjuryOct 5, 2018Migration or Expulsion of Device; Migration or Expulsion of Device

    ADDITIONAL INFORMATION RECEIVED ON 05OCTOBER2018: THE PATIENT WAS EXPERIENCING STIFFNESS FOLLOWING THE FIRST PROCEDURE IN (B)(6).

    Patient problems: Muscular Rigidity; Tissue Damage

    MAUDE report 7939855 ↗
  • MalfunctionJun 15, 2018Migration or Expulsion of Device; Migration or Expulsion of Device

    ADDITIONAL EVENT INFORMATION RECEIVED ON 5/22/2018. NO REPLACEMENTS WERE ORDERED PER THE SALES REP'S REQUEST INSTEAD A TRUNK STOCK ADJUSTMENT HAS BEEN REQUESTED ON BEHALF OF THE REP PER THEIR REQUEST.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7606430 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →