HEALIX ADVANCE PEEK ANCHOR w/PERMATAPE PERMACORD 222363Recalls & reported problems
Medos International Sàrl · PEEK Anchor with (1) 2.5mm Blue PERMATAPE Braided Flat Suture, 38 inches (96 cm) and (1) White/Black, Size 2 (5 metric) PERMACORD Braided Suture, 36 inches (91cm), (1) Utility Shuttle Suture · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886705027835), received between Dec 28, 2020 and Oct 31, 2025 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component1
- Difficult to Advance1
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Foreign Body In Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2025Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE, THE SURGEON PLACED THE HEALIX AV PK 4.5 TAPE BLUE ANCHOR DEVICE INTO THE BONE TO GO PAST THE FIRST SUTURE, UPON INSTANTLY HE NOTICED THAT THE SUTURE WAS NOT LOADED DIRECTLY INTO THE ANCHOR ITSELF, AND IT WAS TECHNICALLY UNLOADED FROM THE ANCHOR PRIOR TO PLACEMENT. IT WAS REPORTED THAT THERE WAS AN UNSPECIFIED SURGICAL DELAY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23437803 ↗MalfunctionSep 8, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE TAPE CAME OUT OF THE 4.5MM HEALIX ADVANCE¿ PEEK ANCHOR WITH PERMATAPE¿ SUTURE, BLUE DEVICE WHEN THE SURGEON PLACED AND WANTED TO CYCLE THE TAPE SUTURE. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15382685 ↗MalfunctionOct 21, 2021Break; Break
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE TIP OF THE ANCHOR ON THE HEALIX AV PK 4.5 TAPE BLUE DEVICE CRACKED WHEN IMPLANTING. ALL FRAGMENTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A DELAY OF TWO MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12673718 ↗InjuryFeb 11, 2021Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED BY THE AFFILIATE IN SWITZERLAND THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON 1/27/2021, IT WAS OBSERVED THAT THE HEALIX AV PK 4.5 TAPE BLUE DEVICE WAS TORN AS IT WAS THREADED THROUGH THE BONE. THE ANCHOR WAS LEFT INSIDE THE PATIENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A 30 MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. THE STATUS OF THE PATIENT POST-SURGERY WAS UNKNOWN. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 11315803 ↗InjuryDec 28, 2020Break; Break
IT WAS REPORTED BY THE SALES REP VIA PHONE THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2020, IT WAS OBSERVED THAT A HEALIX AV PK 4.5 TAPE BLUE DEVICE BROKE DURING INSERTION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11077565 ↗
Other Fastener, fixation, nondegradable, soft tissue devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,686 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,398 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE932 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE721 reports
- RIGIDLOOP 232447Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE633 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue624 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE609 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE460 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.