Skip to content

HEALIX ADVANCE PEEK ANCHOR w/PERMATAPE PERMACORD 222365Recalls & reported problems

Medos International Sàrl · PEEK Anchor with (1) 2.5mm Blue PERMATAPE Braided Flat Suture, 38 inches (96cm) and (1) White/Black, Size 2 (5 metric) PERMACORD Braided Suture, 36 inches (91cm), (1) Utility Shuttle Suture · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886705027996), received between Feb 8, 2021 and Feb 8, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult to Advance1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 8, 2021Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED BY THE AFFILIATE IN AUSTRALIA THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE HEALIX AV PK 5.5 TAPE BLUE ANCHOR DEVICE FAILED TO ADVANCE DUE TO DEFORMATION OF CORTICAL THREAD. ACCORDING TO THE REPORT, THE ANCHOR WAS SPINNING WITHIN THE HOLE AND NOTED THAT THE FIRST CORTICAL THREAD WAS DEFORMED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A TWO MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11293834 ↗

Other Fastener, fixation, nondegradable, soft tissue devices

All Fastener, fixation, nondegradable, soft tissue devices →

Sources and updates