Skip to content

NA 03.010.061SRecalls & reported problems

Synthes GmbH · 4.2MM THREE-FLUTED DRILL BIT QC/330MM/100MM CALIB-STERILE · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982067340), received between Apr 26, 2021 and Jan 27, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device-Device Incompatibility1
  • Entrapment of Device1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 27, 2026Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2026, THE PATIENT UNDERWENT AN UNKNOWN SURGERY WITH THE DRILL BIT. WHEN THE SURGEON WAS TRYING TO LOCK THE DISTAL STENT, HE PRESSED THE DEVICE TOO HARD, CAUSING INTERFERENCE WITH THE NAIL. THE SURGEON USED A HAMMER TO KNOCK THE DEVICE THROUGH THE NAIL HOLE. AFTER CONNECTING THE DRILL TO THE POWER TOOL AND STARTING DRILLING, THE TIP OF DRILL BIT BROKE. THE DRILL TIP REMAINED INSIDE THE BODY, SO IT WAS EXCAVATED FROM THE PROXIMAL CORTICAL BONE AND REMOVED. AFTER THAT, THE SURGEON DRILLED FREEHAND AND INSERTED THE LOCKING SCREW. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHIN 30MINS OF DELAY. THE SURGEON COMMENTED THAT THE CORTICAL BONE WAS THICK, SO IT SEEMS THAT UNUSUAL STRESS WAS GENERATED IN THE DRILL BIT, CAUSING IT TO BREAK. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 24189124 ↗
  • MalfunctionApr 26, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021 WHILE DRILLING FOR THE DISTAL SCREW OF A RETROGRADE/ANTEGRADE FEMORAL NAIL, THROUGH THE SPIRAL BLADE ATTACHMENT, THE DRILL BIT AND SCREW BOTH SKIVED POSTERIORLY MISSING THE NAIL ALL TOGETHER. THE SCREW WAS ABORTED AND THE SPIRAL BLADE WAS COMPLETED WITHOUT INCIDENT. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY REPORTED. THIS REPORT IS FOR ONE (1) 4.2MM THREE-FLUTED DRILL BIT QC/330MM/100MM CALIB-STERILE. THIS IS REPORT 8 OF 10 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11726972 ↗

Other BIT, DRILL devices