NA 03.010.104Recalls & reported problems
Synthes GmbH · 4.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is a Synthes 4.2 mm three-fluted drill bit used to drill into bone during orthopedic surgery. Most of the 54 reports describe the bit tip snapping off during drilling, sometimes leaving a fragment embedded in the bone, along with bits becoming stuck or not cutting as expected. No deaths were reported and 13 reports involved an injury. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 53
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 53 adverse event reports that list this exact device (DI 10886982067746), received between Feb 23, 2017 and Jun 16, 2026 — classified as 13 injuries, 40 malfunctions.
The most frequently reported specific device problem is “Break” (41 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break41
- Entrapment of Device18
- Device-Device Incompatibility5
- Device Operates Differently Than Expected3
- Failure to Cut2
- Contamination /Decontamination Problem1
- Device Damaged Prior to Use1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- Foreign Body In Patient21
- Insufficient Information8
- No Patient Involvement4
- Device Embedded In Tissue or Plaque2
- No Code Available2
- Injury1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 53 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2026Break; Entrapment of Device
IT WAS REPORTED THAT THE TIP OF THE DRILL BIT BREAKS OFF DURING THE PROCEDURE AND REMAINS EMBEDDED IN THE PATIENT'S BONE.
Patient problems: Foreign Body In Patient
MAUDE report 25522750 ↗MalfunctionMay 20, 2026Break
IT WAS REPORTED THAT IN THE SURGERY THE SPECIALIST BROKE THE TIP OF THE 4.2 BIT BY PIERCING THE BONE.
Patient problems: Insufficient Information
MAUDE report 25231030 ↗MalfunctionMay 6, 2026Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE TWO DRILL BITS BREAK WITHOUT AFFECTING THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25092641 ↗MalfunctionMay 5, 2026Break
IT WAS REPORTED THAT DURING THE PROCEDURE, THE TWO DRILL BITS BREAK WITHOUT AFFECTING THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25084462 ↗MalfunctionMay 5, 2026Break
IT WAS REPORTED THAT THE DRILL BIT Ø4.2 CALIBR L145 3FLUTE F/QUIC BROKE AT THE TIP DURING AN INTRAOPERATIVE PROCEDURE. THE EVENT OCCURRED WHEN THE DRILL BIT WAS USED AFTER A POLE WAS PLACED TO DIRECT A NAIL; THE BROKEN FRAGMENT WAS REMOVED AND THE SCREW WAS PLACED IMMEDIATELY. THERE WAS NO REPORTED CONSEQUENCE OR IMPACT TO THE PATIENT, AND NO SIGNS, SYMPTOMS, OR PATIENT INVOLVEMENT WERE IDENTIFIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25084273 ↗MalfunctionApr 6, 2026Break
IT WAS REPORTED THAT DURING THE DISTAL BLOCKAGE OF THE NAIL, THE TIP OF THE BIT WAS FRACTURED; IT WAS SENT WITH RED TAPE. THE CONNECTING SCREW DOES NOT HOLD THE SPIRAL SHEET WELL: THE THREAD STEP FEELS HARD WHEN ATTEMPTING TO THREAD AND DOES NOT COMPLETE THE TURNS NECESSARY TO HOLD THE SHEET PROPERLY. THE IMPACTOR HANDLE OF THE BLADE IS LOOSE IN THE RUBBER, WHICH MAKES IT DIFFICULT TO HOLD ON IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24790393 ↗MalfunctionMar 4, 2026Break
IT WAS REPORTED THE DRILL BIT BROKE DURING THE DRILLING. THERE WERE NO COMPLICATIONS; THE SURGERY WAS SUCCESSFUL. THERE WAS ANOTHER DRILL BIT AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24512171 ↗MalfunctionMar 4, 2026Material Deformation
IT WAS REPORTED THAT EMPHASYS TO PINNACLE CUP ADAPTOR THREADS WERE STRIPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24507148 ↗MalfunctionJan 22, 2026Break
IT WAS REPORTED THE END OF DRILL BIT HAS SNAPPED OFF INSIDE THE PATIENT'S TIBIA. RETRIEVAL OF THE END OF THE DRILL BIT WAS ATTEMPTED BUT WAS UNSUCCESSFUL. AN X-RAY WAS TAKEN AND SAVED AS EVIDENCE OF THE LOCATION OF THE DRILL BIT THAT IS NOT HARMING THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24149274 ↗MalfunctionDec 12, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED TWO (2) 4.2 MM CALIBRATED DRILL BIT, 145 MM LONG, BROKE AT THE TIP WHILE BEING USED TO DRILL DISTAL LOCKING DEVICES. TWO FRAGMENTS REMAINED EMBEDDED IN THE PATIENT'S BONE. A SHORT RETAINING PIN FOR A SCREWDRIVER ALSO BROKE AT THE TIP, LEAVING FRAGMENTS EMBEDDED IN THE SCREW, WHICH IS ALSO IN THE PATIENT'S BONE.
Patient problems: Foreign Body In Patient
MAUDE report 23788303 ↗MalfunctionDec 12, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED TWO (2) 4.2 MM CALIBRATED DRILL BIT, 145 MM LONG, BROKE AT THE TIP WHILE BEING USED TO DRILL DISTAL LOCKING DEVICES. TWO FRAGMENTS REMAINED EMBEDDED IN THE PATIENT'S BONE. A SHORT RETAINING PIN FOR A SCREWDRIVER ALSO BROKE AT THE TIP, LEAVING FRAGMENTS EMBEDDED IN THE SCREW, WHICH IS ALSO IN THE PATIENT'S BONE.
Patient problems: Foreign Body In Patient
MAUDE report 23788219 ↗MalfunctionJul 10, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT DURING THE PROCEDURE OF RETROGRADE FEMORAL RODDING WITH RETROGRADE FEMORAL NAIL ADVANCED, IT WAS REPORTED THAT, PRIOR TO INSERTION OF RETROGRADE NAIL, SURGEON USED A 4.2MM NEEDLE POINT DRILL (CATALOG NUMBER 03.010.104) AS 'BLOCKING SCREW' TECHNIQUE OR 'POLLER SCREW' AS A REDUCTION AID. DRILL PERFORMED AS INTENDED. AT END OF PROCEDURE, AFTER ALL IMPLANTS WERE IN, THE AFOREMENTIONED DRILL WAS REMOVED, AND THE DISTAL 1CM OF DRILL FRACTURED FROM THE BASED OF THE DRILL AND REMAINED IN THE PATIENTS FEMUR. THE SURGEON DECIDED TO LEAVE THE FRAGMENT OF DRILL AS THE IMPLANT CONSTRUCT WAS UNAFFECTED AND RETRIEVING THE BROKEN DRILL PIECE WAS DETERMINED TO DO MORE HARM THAN JUST LEAVING IT. THERE WAS NO SURGICAL DELAY. PROCEDURE COMPLETED SUCCESSFULLY.
Patient problems: Foreign Body In Patient
MAUDE report 22474079 ↗MalfunctionJun 24, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THE TFNA AIMING ARM AND ASSOCIATED PIECES GET JAMMED AND STUCK BEFORE FULL INSERTION IS ACHIEVED; IT WAS NOTED ABNORMALLY HIGH RESISTANCE IS EXPERIENCED WHEN INSERTING GUIDE SLEEVES. THIS WAS DISCOVERED DURING REPROCESSING. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22321993 ↗MalfunctionJun 24, 2025Device-Device Incompatibility; Device-Device Incompatibility
TFNA AIMING ARM AND ALL ASSOCIATED PIECES NEED TO BE REPLACED DUE TO DIFFICULTY AND ABNORMALLY HIGH RESISTANCE EXPERIENCED WHEN INSERTING GUIDE SLEEVES. THEY APPEARED TO GET JAMMED AND STUCK BEFORE FULL INSERTION WAS ACHIEVED. THIS WAS DISCOVERED DURING REPROCESSING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22320844 ↗MalfunctionJun 16, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED DURING AN ADVANCED TIBIA NAIL PROCEDURE, THE DRILL PLUNGED OFF ANGLE AND CAUGHT THE HOLE IN THE NAIL AND BROKE. NO SURGICAL DELAY. PROCEDURE WAS COMPLETED SUCCESSFULLY. IT WAS COMPLETED BY DRILLING ANOTHER HOLE. FRAGMENTS WERE GENERATED AND THE DRILL BIT TIP WAS LEFT IN LEG,
Patient problems: Foreign Body In Patient
MAUDE report 22211170 ↗MalfunctionApr 30, 2025Break; Break
IT WAS REPORTED THAT THE DOCTOR FOLDS THE TIP OF THE GUIDE SO THAT IT CAN PASS INTO THE SPINAL CANAL. THE DOCTOR BEGINS TO MAKE PERFORATIONS IN THE BONE VERY HARD AND THE FLANGE THAT BREAKS IN THE MIDDLE OF THE SURGERY. THE DOCTOR MAKES A LOT OF FORCE ON THE HEAD OF THE SCREW WITH THE SCREWDRIVER AND THE TIP OF THE SCREW IS BROKEN.
Patient problems: Insufficient Information
MAUDE report 21936118 ↗MalfunctionApr 23, 2025Break; Break
IT BENDS MANUAL MEDULLAR FRESE #6 AND #8 ALSO BREAK A 4.2 L145 BIT AND ANOTHER 4.2 L340 BIT. ALSO MANGO IN T DURING THE SURGERY THE COMPANION OF CENTRAL WASHED AREA IS INFORMED.
Patient problems: Insufficient Information
MAUDE report 21892308 ↗MalfunctionMar 6, 2025Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT DURING A RETROGRADE FEMORAL NAIL PROCEDURE, THE 4.2 MM DRILL BIT BROKE OFF IN PATIENT AND WAS NOT RETRIEVED.
Patient problems: Foreign Body In Patient
MAUDE report 21535313 ↗MalfunctionMay 13, 2024Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS NOTED ON APRIL 1, 2024, PRE-OPERATIVELY, A 4.2MM CALIBER DRILL BIT'S TIP IS MISSING. THIS REPORT IS FOR ONE 4.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19297686 ↗MalfunctionMar 25, 2024Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT ON AN UNKNOWN DATE DRILL BUT TIP BROKE OFF DRILLING FOR DISTAL SCREWS USING PERFECT CIRCLE TECHNIQUE. SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. PATIENT OUTCOME WAS UNKNOWN. FRAGMENTS WERE GENERATED. THIS REPORT IS FOR ONE (1) 4.2 3-FLTD DRILL BIT QC/NEEDLE PT/145 THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 18969521 ↗MalfunctionFeb 20, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2023, THAT A LOAN TFNA SET WAS FOUND TO HAVE BONE CONTAMINANTS IN THE DRILL BITS DURING CHECKING IN BY CSSD. PATIENT STATUS/ OUTCOME / CONSEQUENCES :NO. THIS REPORT IS FOR ONE (1) UNK - DRILL BITS: TRAUMA. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18741001 ↗MalfunctionFeb 2, 2024Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2024 WHILE PERFORMING DISTAL BLOCKAGES, WHEN MANUALLY PASSING THE DRILL BIT IN QUESTION, THE BIT BROKE. A PART OF THE DRILL BIT REMAINED IN THE PATIENT; IT COULD NOT BE REMOVED SINCE IT WAS INSIDE OF THE NAIL HOLE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 4.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 18630847 ↗MalfunctionJan 10, 2024Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED THAT ON AN UNKNOWN DATE, A DRILL BIT BROKE DURING A FEMORAL RECON NAIL PROCEDURE. THE TIP OF THE BROKEN DRILL TIP WAS LEFT IN THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE 4.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 18482526 ↗MalfunctionNov 30, 2023Break; Break
DEVICE REPORT FROM THE UNITED KINGDOM REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON 08-NOVEMBER-2023, THE DRILL BIT IN QUESTION BROKE INTRA-OP. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY USING A REPLACEMENT DRILL BIT OF THE SAME TYPE. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 4.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 18241935 ↗MalfunctionNov 8, 2023Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE DRILL BIT IN QUESTION BROKE WHILE STARTING DISTAL LOCKING. THERE WAS A FRAGMENT GENERATED THAT REMAINS IN THE PATIENT. THE DRILL BIT WAS CHANGED, AND THE PROCEDURE WAS ABLE TO BE COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 4.2MM THREE-FLUTED DRILL BIT QC/NEEDLE POINT/145MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 18091515 ↗
FDA clearance and classification
The FDA regulates product code HTW (bit, drill) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 03.010.104
Has the NA 03.010.104 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982067746), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 03.010.104?
The FDA's MAUDE database lists 53 adverse event reports for this device, received between Feb 23, 2017 and Jun 16, 2026. In those reports, the most common specific device problems are “Break” (41), “Entrapment of Device” (18) and “Device-Device Incompatibility” (5); the most common patient problems are “Foreign Body In Patient” (21), “Device Embedded In Tissue or Plaque” (2) and “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 03.010.104?
None of the reports are classified as deaths, 13 as injuries and 40 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 03.010.104 FDA approved?
Devices of this type (product code HTW) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the NA 03.010.104, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 03.010.104). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,978 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL769 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL335 reports
- NA 323.062Synthes GmbH · BIT, DRILL308 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL210 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL204 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026