Skip to content

NA 03.100.109Recalls & reported problems

Synthes GmbH · ALUMINUM HOHMANN RETRACTOR 18MM WIDTH · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886982071071), received between Jan 10, 2018 and Feb 19, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Material Deformation2
  • Crack1
  • Device-Device Incompatibility1
  • Material Fragmentation1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2026Material Integrity Problem

    IT WAS REPORTED THE ISSUE HAS BEEN IDENTIFIED DURING LOAN KIT INSPECTION, BY THE LOAN KIT TECHNICIAN. THE EVENT DATE AND ALERT DATE ARE THE DATE THAT THE INSPECTION TOOK PLACE. THERE IS NO SURGEON, PROCEDURE, OR PATIENT DETAILS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24388933 ↗
  • MalfunctionDec 15, 2025Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT AN UNKNOWN SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. A DAY AFTER THE SURGERY ON (B)(6) 2025, THE SALES REP WAS INFORMED THAT A NURSE DISCOVERED THAT THE TIP OF THE RETRACTOR APPEARED TO HAVE A SMALL CHIP BEFORE AND AFTER CLEANING THE INSTRUMENT. IT IS UNKNOWN WHETHER ANY REMAINED IN THE PATIENT'S BODY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23809455 ↗
  • MalfunctionFeb 2, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT ON DECEMBER 8, 2023, IT WAS NOTICED THAT THERE WERE SOME DENTS AND DAMAGES TO THE INSTRUMENTS. THERE WAS NO PATIENT INVOLVEMENT; THIS WAS NOTED UPON INSPECTION OF THE TRAYS. NO SURGICAL CASES WERE AFFECTED. UPON INSPECTION OF THE RETURNED DEVICES, IT WAS NOTED THAT THE SDDRIVE SCREWDRIVER SHAFT T8 105MM WAS STRIPPED AND DEFORMED, THE HOHMANN RETRACTOR 15MM 160MM WAS WORN AND DEFORMED, THE ALUMINUM HOHMANN RETRACTOR 18MM WIDTH WAS BROKEN, AND THE ALUMINUM HOHMANN RETRACTOR 18MM WIDTH WAS DEFORMED AND WORN. THIS REPORT INVOLVES ONE ALUMINUM HOHMANN RETRACTOR 18MM WIDTH. THIS IS REPORT 4 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18633300 ↗
  • MalfunctionFeb 2, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT ON DECEMBER 8, 2023, IT WAS NOTICED THAT THERE WERE SOME DENTS AND DAMAGES TO THE INSTRUMENTS. THERE WAS NO PATIENT INVOLVEMENT; THIS WAS NOTED UPON INSPECTION OF THE TRAYS. NO SURGICAL CASES WERE AFFECTED. UPON INSPECTION OF THE RETURNED DEVICES, IT WAS NOTED THAT THE SDDRIVE SCREWDRIVER SHAFT T8 105MM WAS STRIPPED AND DEFORMED, THE HOHMANN RETRACTOR 15MM 160MM WAS WORN AND DEFORMED, THE ALUMINUM HOHMANN RETRACTOR 18MM WIDTH WAS BROKEN, AND THE ALUMINUM HOHMANN RETRACTOR 18MM WIDTH WAS DEFORMED AND WORN. THIS REPORT INVOLVES ONE ALUMINUM HOHMANN RETRACTOR 18MM WIDTH. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18633267 ↗
  • MalfunctionMay 31, 2023Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2023, DURING A SURGERY THE TIP OF A RADIOLUCENT HOHMANN RETRACTOR WAS BROKE OFF. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THE BROKEN TIP WAS RETRIEVED WITHOUT INCIDENT. THIS REPORT IS FOR ONE (1) ALUMINUM HOHMANN RETRACTOR 18MM WIDTH. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17033273 ↗
  • MalfunctionMar 21, 2023Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: THIS REPORT IS FOR ONE (1) RADIOLUCENT HOHMANN RETRACTOR W/18 L240.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16583206 ↗
  • MalfunctionDec 6, 2021Crack; Device-Device Incompatibility; Crack; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, UNABLE TO STERILIZE/ASSEMBLE THE MALLEABLE RETRACTOR. CHIPPING AND CRACKING ON INSTRUMENTS. UNABLE TO USE. THIS COMPLAINT INVOLVES TEN(10) DEVICES. THIS REPORT IS FOR (1) ALUMINUM HOHMANN RETRACTOR 18MM WIDTH. THIS IS REPORT 10 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12938881 ↗
  • MalfunctionFeb 13, 2020Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE SURGERY FOR PLATEAU FRACTURE WAS PERFORMED WITH RADIOLUCENT HOHMANN RETRACTOR. DURING THE SURGERY, A NURSE FOUND THE TIP OF THE RETRACTOR HAD BROKEN. THE SURGEON THOUGHT THAT IT BROKE DURING DRILLING. NO FURTHER INFORMATION IS AVAILABLE. CONCOMITANT DEVICE REPORTED: UNK - DRILL (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY 1). THIS REPORT IS FOR ONE (1) ALUMINUM HOHMANN RETRACTOR 18MM WIDTH. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9706543 ↗
  • MalfunctionJan 10, 2018Break; Material Fragmentation; Break; Material Fragmentation

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THE ALUMINUM HOHMANN RETRACTOR WAS USED DURING A PROCEDURE TO REPAIR A FEMORAL TROCHANTERIC FRACTURE ON (B)(6) 2017. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. AFTER THE PROCEDURE THE STAFF WAS WASHING AND INSPECTING THE DEVICE AND NOTICED THAT THE TIP OF THE RETRACTOR WAS BROKEN. IT IS NOT KNOWN IF THE DEVICE BROKE INTRAOPERATIVELY OR POSTOPERATIVELY. NO CONFIRMATION WAS MADE TO DETERMINE IF THE BROKEN TIP REMAINS IN THE PATIENT. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7178379 ↗

Other RETRACTOR devices