NA 03.114.007Recalls & reported problems
Synthes GmbH · 1.1MM DRILL BIT/MQC/75MM · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10886982073013), received between Nov 14, 2022 and Dec 30, 2025 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Cut11
- Material Deformation3
- Break1
- Device-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information6
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THE DRILL BITS WERE DULL WHEN DRILLING THE CHANNEL. ANOTHER DRILL BIT WAS USED TO COMPLETE THE PROCEDURE WITH NO DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23921660 ↗MalfunctionApr 24, 2025Failure to Cut; Failure to Cut
IT WAS REPORTED THE 1.1 DRILL BIT DID NOT HAVE ENOUGH CUTTING POWER AND REQUIRED EXCESSIVE FORCE TO WORK. THIS HAPPENED WHEN THE EQUIPMENT WAS OPENED.
Patient problems: Insufficient Information
MAUDE report 21902775 ↗MalfunctionMay 3, 2024Failure to Cut; Material Deformation; Failure to Cut; Material Deformation
IT WAS REPORTED DURING A SURGERY ON (B)(6) 2024, THE DRILL BITS TWISTED DURING THE PROCEDURE AND HAD NO EDGE. SURGERY WAS COMPLETED SUCCESSFULLY WITH AN UNKNOWN SURGICAL DELAY. THIS REPORT IS FOR ONE 1.1MM DRILL BIT/MQC/75MM FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 19239960 ↗MalfunctionMay 3, 2024Failure to Cut; Material Deformation; Failure to Cut; Material Deformation
IT WAS REPORTED DURING A SURGERY ON (B)(6) 2024, THE DRILL BITS TWISTED DURING THE PROCEDURE AND HAD NO EDGE. SURGERY WAS COMPLETED SUCCESSFULLY WITH AN UNKNOWN SURGICAL DELAY. THIS REPORT IS FOR ONE 1.1MM DRILL BIT/MQC/75MM FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 19239844 ↗MalfunctionDec 1, 2023Failure to Cut; Failure to Cut
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, THE DRILL BITS IN QUESTION WERE FOUND TO BE DULL AND WOULD NOT CUT AS INTENDED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 10 MINUTES. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 18246179 ↗MalfunctionDec 1, 2023Failure to Cut; Failure to Cut
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, THE DRILL BITS IN QUESTION WERE FOUND TO BE DULL AND WOULD NOT CUT AS INTENDED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF 10 MINUTES. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 18246178 ↗MalfunctionNov 20, 2023Failure to Cut; Material Deformation; Failure to Cut; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2023, THAT THE DRILL BIT 1.1 L75/61 2 ARIST CUTTING DRILL BITS TWISTED SPECIALIST NON-CONFORMING. DRILL BIT 1.0 L46/34, BROKEN DRILL BITS, SPECIALIST NON-CONFORMING GUIDE DRILL BIT 1.3/1.0, GUIDE COMES IN TWISTED SYSTEM BY SIDE 1.0, DOES NOT PASS DRILL BIT 1.0. PROLONGATION OF SURGICAL TIMES. THIS REPORT IS FOR ONE (1) 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 1 OF 4 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18173201 ↗MalfunctionMay 22, 2023Failure to Cut; Failure to Cut
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, BOTH Ø1.1 BITS FROM THE 1.5 SYSTEM WERE FOUND DAMAGED WITH NO CUTTING EDGE. NO PATIENT IMPACT WAS REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16974753 ↗MalfunctionMay 22, 2023Failure to Cut; Failure to Cut
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6)2023, BOTH Ø1.1 BITS FROM THE 1.5 SYSTEM WERE FOUND DAMAGED WITH NO CUTTING EDGE. NO PATIENT IMPACT WAS REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 1 OF 2 FOR (B)(4)
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16974748 ↗MalfunctionMay 22, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS DISCOVERED ON (B)(6) 2023 THAT THE DRILL BIT IN QUESTION WOULD NOT PROPERLY ANCHOR INTO THE MOTOR ADAPTER. THERE WAS NO REPORTED PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16973378 ↗MalfunctionFeb 13, 2023Break; Break
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE 1.1 DRILL BIT IN QUESTION FROM THE HAN LCP SYSTEM SPLIT DURING USE. THE PROCEDURE WAS COMPLETED USING ANOTHER DRILL BIT OF THE SAME TYPE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16363008 ↗MalfunctionNov 14, 2022Failure to Cut; Failure to Cut
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, THE SURGEON EXPERIENCED DIFFICULTY DRILLING WITH THE DRILL BITS IN QUESTION. CUTTING WAS NOT GOOD, AND THE DRILLING WAS DELAYED BECAUSE OF IT. THE BIT WAS CHANGED AND THE HOLES WERE ABLE TO BE DRILLED, BUT WITH GREAT DIFFICULTY. ADDITIONALLY, THE GUIDE WOULD NOT SCREW TO THE PLATES OF 1.5, IT COULD NOT BE USED. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15787845 ↗MalfunctionNov 14, 2022Failure to Cut; Failure to Cut
DEVICE REPORT FROM COLOMBIA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, THE SURGEON EXPERIENCED DIFFICULTY DRILLING WITH THE DRILL BITS IN QUESTION. CUTTING WAS NOT GOOD, AND THE DRILLING WAS DELAYED BECAUSE OF IT. THE BIT WAS CHANGED AND THE HOLES WERE ABLE TO BE DRILLED, BUT WITH GREAT DIFFICULTY. ADDITIONALLY, THE GUIDE WOULD NOT SCREW TO THE PLATES OF 1.5, IT COULD NOT BE USED. THIS REPORT IS FOR A 1.1MM DRILL BIT/MQC/75MM. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15787844 ↗
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,985 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL770 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL339 reports
- NA 323.062Synthes GmbH · BIT, DRILL309 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL211 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL207 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.