NA 03.130.100Recalls & reported problems
Synthes GmbH · 1.0MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/75MM · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10886982075215), received between Apr 5, 2017 and Mar 25, 2026 — classified as 5 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Break” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break10
- Entrapment of Device2
- Failure to Cut2
- Material Fragmentation2
Patient problems
- Foreign Body In Patient5
- No Clinical Signs, Symptoms or Conditions5
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2026Break
IT WAS REPORTED DURING THE PROCEDURE, WHEN THE SPECIALIST MADE THE BROCADE WITH ONE OF THE DRILL BITS FOR CORE HOLE AND FOR THREE-JAW MANDREL, IT BROKE. A PROMPT SOLUTION WAS GIVEN: THE BITS OF THE BIT WERE REMOVED AND A NEW BIT OF THE SAME REFERENCE AND CHARACTERISTICS WAS GIVEN TO DR. WITH THIS BIT IT WAS POSSIBLE TO COMPLETE THE SURGERY SUCCESSFULLY, WITHOUT AFFECTING THE PATIENT AND WITHOUT ALTERATIONS IN THE SURGICAL TIMES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24690923 ↗MalfunctionMar 23, 2026Failure to Cut
IT WAS REPORTED DURING SURGERY, SURGEON COMPLAINED THAT THE IDENTIFIED DRILL BITS WERE DULL AND WERE NO LONGER FUNCTIONING TO THEIR OPTIMAL DRILLING CAPACITY. THESE COMPONENTS WERE DEEMED TO BE NO LONGER FUNCTIONAL AND WERE SUBSEQUENTLY DISCARDED. IN ADDITION, MEDICAL DEVICE REPROCESSING STAFF IDENTIFIED THAT THE TIPS OF THE COMPLAINED SCREWDRIVER SHAFTS WERE BENT AND SLIGHTLY DEFORMED. THESE ITEMS WERE DEEMED TO BE NO LONGER FUNCTIONAL AND WERE SUBSEQUENTLY DISCARDED. NO SURGICAL DELAY. NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660338 ↗MalfunctionMar 18, 2026Failure to Cut
DURING SURGERY, IT WAS NOTED THAT THE IDENTIFIED DRILL BITS LISTED IN THIS COMPLAINT WERE DULL AND WERE NO LONGER FUNCTIONING TO THEIR OPTIMAL DRILLING CAPACITY. THESE COMPONENTS WERE DEEMED TO BE NO LONGER FUNCTIONAL, AND WERE SUBSEQUENTLY DISCARDED. REPLACEMENTS ARE REQUIRED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24634709 ↗MalfunctionSep 19, 2024Break; Entrapment of Device; Break; Entrapment of Device
DURING THE PROCEDURE ON (B)(6) 2024, AT THE TIME THE DRILLING, THE DRILL BIT BROKE AND A SMALL FRAGMENT OF IT REMAINED INSIDE THE BONE OF THE PATIENT. THE SURGEON TRIED TO REMOVE THIS SMALL FRAGMENT OF THE DRILL BIT, BUT IT WAS NOT POSSIBLE. THE SURGEON EVALUATED THE AREA WHERE THE FRAGMENT IS LEFT AND DECIDED TO CONTINUE WITH THE PROCEDURE WITH A NEW DRILL BIT. THE SURGERY WAS COMPLETED SUCCESSFULLY, WITHOUT ANY ADDITIONAL AFFECTATION TO THE PATIENT AND WITHOUT ALTERATIONS IN THE SURGICAL TIMES. THE PATIENT'S CONDITION DURING AND AFTER THE SURGERY WAS STABLE AND WITHOUT ADDITIONAL COMPLICATIONS.
Patient problems: Foreign Body In Patient
MAUDE report 20260499 ↗MalfunctionMay 16, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, WHILE USING THE DRILL BIT TO OPEN A CHANNEL IN THE JOINT, THE DRILL BIT BROKE. FRAGMENTS WERE GENERATED, BUT THESE WERE REMOVED WITHOUT ANY PROBLEM AND THERE WAS NO NEED FOR ADDITIONAL INTERVENTION FOR THE PATIENT. ANOTHER DRILL BIT WAS USED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO SURGICAL DELAY AND NO PATIENT HARM. THIS REPORT IS FOR ONE (1) 1.0MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/75MM THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19335781 ↗MalfunctionAug 23, 2023Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLUMBIA AS FOLLOWS: IT WAS REPORTED DURING SURGERY ON (B)(6) 2023, THE 1.0 DRILL BIT BROKE. THIS REPORT IS FOR ONE 1.0MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/75MM FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 17605441 ↗InjuryApr 1, 2022Entrapment of Device; Entrapment of Device
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE UNKNOWN DRILL BIT BROKE. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS REPORT IS FOR ONE (1) 1.0 DRILL BIT/K-WIRE ATTCHMT THRD H/75. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 13976240 ↗MalfunctionMar 17, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT AN UNKNOWN PROCEDURE USING VARIABLE ANGLE HANDSET. DURING THE PROCEDURE, WHILE SURGEON WAS USING A DRILL BIT THROUGH A LOCKING PLATE (VA HANDSET) ON A PROXIMAL PHALANXE, THE DRILL BIT BROKE. THE FRAGMENT WAS REMOVED. THE SURGEON RETRIEVED THE FRAGMENT AND THE SURGERY WAS COMPLETED SUCCESSFULLY WITH A FIVE (5) MINUTE SURGICAL DELAY. THE PATIENT'S STATUS IS UNKNOWN. CONCOMITANT DEVICE REPORTED: UNKNOWN 1.3 LOCKING PLATE (PART#: UNKNOWN, LOT#: UNKNOWN, QUANTITY: 1) THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) 1.0MM DRILL BIT/K-WIRE ATTCHMT FOR THREADED HOLE/75MM THIS REPORT IS 1 OF 1 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11508520 ↗MalfunctionOct 6, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING THE SURGICAL PROCEDURE ON (B)(6) 2020, ONE (1) 1.0MM DRILL BIT WAS BROKEN AND ONE (1) 1.0MM DRILL BIT WAS BENT. NO FRAGMENTS REMAINED IN THE PATIENT. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO OTHER MEDICAL INTERVENTION. NO FURTHER INFORMATION PROVIDED. CONCOMITANT DEVICE REPORTED: LOCK STRUT PL 1.3 8HO OBLIQ-ANGL LE L19 (PART # 02.130.159, LOT # L483064, QUANTITY 1). THIS REPORT IS FOR ONE (1) 1.0MM DRILL BIT/ K-WIRE ATTCHMT FOR THREADED HOLE/ 75MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10638931 ↗InjuryMar 14, 2020Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT ON (B)(6) 2020, THE SURGEON WAS DRILLING WITH A 1.0 DRILL BIT AND AFTER GOING THROUGH THE FIRST CORTEX, THE DRILL BIT BROKE BEFORE PENETRATING THE SECOND CORTEX. THE TIP OF THE FLUTED DRILL BIT APPROXIMATELY 5 MM WAS RETAINED IN THE PATIENT. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE REPORTED: UNKNOWN DRILL (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 9832889 ↗InjuryJan 12, 2018Break; Break
IT WAS REPORTED THAT THE PATIENT HAD ORIGINAL SURGERY ON (B)(6) 2017 FOR TREATMENT OF A THIRD METACARPAL FRACTURE. INTER-OPERATIVE, AS THE SURGEON WAS DRILLING A PILOT HOLE THRU THE 1.3MM LOCKING T -PLATE, THE TIP OF THE 1.0MM DRILL BIT BROKE OFF AND BECAME EMBEDDED IN THE PATIENT¿S METACARPAL BONE. SURGEON THEN ATTEMPTED TO IMPLANT A 1.3MM LOCKING SCREW BUT THE SCREW HEAD BROKE OFF DUE TO THE DISTAL PORTION OF THE SCREW MET THE TIP OF THE PREVIOUSLY EMBEDDED DRILL BIT. THE DISTAL PORTION OF THE 1.3MM LOCKING SCREW REMAINS IN THE PATIENT¿S METACARPAL BONE. SURGEON THEN ATTEMPTED TO DRILL ANOTHER PILOT HOLE, THRU ANOTHER SCREW HOLE POSITION, ON THE LOCKING T-PLATE PLATE BY USING ANOTHER 1.0MM DRILL BIT. AGAIN, THE TIP OF THIS 1.0MM DRILL BIT BROKE OFF AND BECAME IMBEDDED IN THE PATIENT¿S METACARPAL BONE. MOVING FORWARDED ON THE PROCEDURE, SURGEON DECIDED TO REMOVE THE LOCKING T-PLATE. SURGEON SELECTED A 12-HOLE LOCKING PLATE AND SCREWS TO COMPRESS THE FRACTURE. UPON COMPLETING THIS PROCEDURE, THE PATIENT WAS ULTIMATELY IMPLANTED WITH ONE (1) 1.3MM LOCKING 12-HOLE STRAIGHT HOLE PLATE AND A COMBINATION LENGTH OF SIX (6) 1.3MM CORTEX STARDRIVE SCREWS. ADDITIONAL X-RAYS WERE TAKEN TO VE
Patient problems: Foreign Body In Patient
MAUDE report 7185893 ↗InjuryJan 12, 2018Break; Break
IT WAS REPORTED THAT THE PATIENT HAD ORIGINAL SURGERY ON (B)(6) 2017 FOR TREATMENT OF A THIRD METACARPAL FRACTURE. INTER-OPERATIVE, AS THE SURGEON WAS DRILLING A PILOT HOLE THRU THE 1.3MM LOCKING T -PLATE, THE TIP OF THE 1.0MM DRILL BIT BROKE OFF AND BECAME EMBEDDED IN THE PATIENT¿S METACARPAL BONE. SURGEON THEN ATTEMPTED TO IMPLANT A 1.3MM LOCKING SCREW BUT THE SCREW HEAD BROKE OFF DUE TO THE DISTAL PORTION OF THE SCREW MET THE TIP OF THE PREVIOUSLY EMBEDDED DRILL BIT. THE DISTAL PORTION OF THE 1.3MM LOCKING SCREW REMAINS IN THE PATIENT¿S METACARPAL BONE. SURGEON THEN ATTEMPTED TO DRILL ANOTHER PILOT HOLE, THRU ANOTHER SCREW HOLE POSITION, ON THE LOCKING T-PLATE PLATE BY USING ANOTHER 1.0MM DRILL BIT. AGAIN, THE TIP OF THIS 1.0MM DRILL BIT BROKE OFF AND BECAME IMBEDDED IN THE PATIENT¿S METACARPAL BONE. MOVING FORWARDED ON THE PROCEDURE, SURGEON DECIDED TO REMOVE THE LOCKING T-PLATE. SURGEON SELECTED A 12-HOLE LOCKING PLATE AND SCREWS TO COMPRESS THE FRACTURE. UPON COMPLETING THIS PROCEDURE, THE PATIENT WAS ULTIMATELY IMPLANTED WITH ONE (1) 1.3MM LOCKING 12-HOLE STRAIGHT HOLE PLATE AND A COMBINATION LENGTH OF SIX (6) 1.3MM CORTEX STARDRIVE SCREWS. ADDITIONAL X-RAYS WERE TAKEN TO VE
Patient problems: Foreign Body In Patient
MAUDE report 7185843 ↗InjuryApr 5, 2017Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT DURING AN OPEN REDUCTION INTERNAL FIXATION OF THE FIRST METACARPAL A 1.0 MM DRILL BIT FROM A MODULAR HAND SYSTEM SHEARED OFF INSIDE THE PATIENT¿S FINGER. THE FRAGMENT WAS RETAINED AND WAS UNABLE TO BE REMOVED. X-RAY TAKEN (B)(6) 2017. SURGERY WAS DELAYED 3-5 MINUTES. PROCEDURE WAS COMPLETED SUCCESSFULLY AND PATIENT OUTCOME WAS GOOD. THIS REPORT IS FOR ONE (1) DRILL BIT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6460045 ↗
Other BIT, DRILL devices
- QUICKSET 2274-25-500Depuy Orthopaedics, Inc. · BIT, DRILL1,985 reports
- QUICKSET 2274-35-500Depuy Orthopaedics, Inc. · BIT, DRILL770 reports
- QUICKSET 2274-30-500Depuy Orthopaedics, Inc. · BIT, DRILL761 reports
- NA 03.037.022Synthes GmbH · BIT, DRILL339 reports
- NA 323.062Synthes GmbH · BIT, DRILL309 reports
- QUICKSET 2274-56-000Depuy (Ireland) · BIT, DRILL275 reports
- QUICKSET 2274-45-500Depuy Orthopaedics, Inc. · BIT, DRILL211 reports
- LCS 2366-83-000Depuy (Ireland) · BIT, DRILL207 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page