Skip to content

NA 03.226.042Recalls & reported problems

Synthes GmbH · 4.0MM CANNULATED DRILL BIT · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982077189), received between Mar 31, 2022 and Nov 11, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem2
  • Break1
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT AN UNKNOWN SURGERY. THE DAY BEFORE THE SURGERY, THE OPERATING ROOM POINTED OUT THAT A TOTAL OF EIGHT ITEMS IN FIVE INSTRUMENT SETS HAD NOT BEEN PROPERLY CLEANED WHEN THEY DELIVERED. ALL THE IMPACTED PRODUCTS OF THE HOLLOW PARTS WERE DIRTY AND APPEARED TO BE BLOOD STAINS. WHEN A BRUSH WAS PASSED THROUGH THE HOLLOW PARTS IN THE OPERATING ROOM, THE BRUSH WAS FOUND TO BE DIRTY, AND SOME INSTRUMENTS WERE SO CLOGGED THAT A BRUSH COULD NOT BE PASSED THROUGH. THE SALES REP ORDERED NEW INSTRUMENTS, HOWEVER MANY OF THEM HAD DROPS OF WATER IN THE HOLLOW PART. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23525835 ↗
  • MalfunctionJul 22, 2024Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT ON (B)(6) 2024, THE PROXIMAL TIBIA SCREWDRIVER OF 3.5 WAS DAMAGED DURING SURGERY, JUST LIKE THE PERFORATION DRILL BIT PREVIOUS. THE THREADED GUIDES ARE STUCK IN THE BONE, AND COULD NOT BE REMOVED. THIS REPORT INVOLVES ONE (1)4.0MM CANNULATED DRILL BIT. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 19797776 ↗
  • MalfunctionMar 31, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION SURGERY. BEFORE SURGERY, AN OPERATION STAFF CHECKED THE PRODUCTS IN QUESTION AND FOUND THAT THERE WERE DIRT/WATER DROPS-LIKE-RESIDUES IN THE HOLLOW PART OF EACH PRODUCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13958646 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page