Skip to content

NA 03.501.040Recalls & reported problems

Synthes GmbH · 2.2MM DRILL BIT STRYKER J-LATCH/WITH 10MM STOP/135MM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10887587010670), received between Oct 11, 2022 and Jan 23, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Failure to Cut2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2024Failure to Cut; Failure to Cut

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLUMBIA AS FOLLOWS: IT WAS REPORTED THAT ON JANUARY 5 2024, THE PRODUCTS HAD NO EDGE. THE EVENT DIDN'T CAUSE ANY HARM TO THE PATIENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) 2.2 DB STRYKER J-LATCH/ 10 STOP/135 THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18566599 ↗
  • MalfunctionJan 23, 2024Failure to Cut; Failure to Cut

    THERE WAS NO SURGICAL DELAY. NO OTHER MEDICAL INTERVENTION WAS REQUIRED. THE SURGERY WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18566082 ↗
  • MalfunctionOct 11, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE DRILL BITS STICK IN CONTACT WITH THE BONE AND BLOCK THE SYSTEM. IT NEGATIVELY AFFECTS THE PROCEDURE DUE TO INCREASED SURGICAL TIME. THERE WERE ALSO TWO DRILL BITS WITHOUT FIXED. THIS REPORT INVOLVES ONE (1) 2.2MM DRILL BIT STRYKER J-LATCH/WITH 10MM STOP/135MM. THIS IS REPORT 6 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15580976 ↗
  • MalfunctionOct 11, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE DRILL BITS STICK IN CONTACT WITH THE BONE AND BLOCK THE SYSTEM. IT NEGATIVELY AFFECTS THE PROCEDURE DUE TO INCREASED SURGICAL TIME. THERE WERE ALSO TWO DRILL BITS WITHOUT FIXED. THIS REPORT INVOLVES ONE (1) 2.2MM DRILL BIT STRYKER J-LATCH/WITH 10MM STOP/135MM. THIS IS REPORT 5 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15580964 ↗

Other BIT, DRILL devices

All BIT, DRILL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page