Medline CPRM2216MRecalls & reported problems
Medline Industries, Inc. · RESUSCITATOR,MAN,PED,MASK,BAG RES,MANO · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Oct 11, 2023Open, Classified
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Z-0369-2024 · Root cause: Device Design · FDA record ↗
Other Ventilator, emergency, manual (resuscitator) devices
- Ventlab AF5140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall8 reports
- Curaplex AF2000Bound Tree Medical, LLC · Ventilator, emergency, manual (resuscitator)1 recall6 reports
- Ventlab AF5140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall5 reports
- Ventlab AF5142MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab AF2140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab AF1140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Ventlab AF2142MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 death1 report
- Ventlab VN3100OBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.