Skip to content

Medline CPRM4416Recalls & reported problems

Medline Industries, Inc. · RESUSCITATOR, SMALL ADLT,MASK,BAG RES · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10888277700482), received between Sep 12, 2026 and Sep 18, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Disconnection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection1
  • Fitting Problem1
  • Inadequate Filtration Process1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Low Oxygen Saturation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Fitting Problem; Material Deformation

    It was reported that "Trying to attach the mask to the bag and could not because the connector was misshapen and would not fit. The connector is hard plastic and could not be manipulated".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50818160 ↗
  • MalfunctionSep 16, 2026Inadequate Filtration Process

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50786943 ↗
  • MalfunctionSep 12, 2026Disconnection

    Patient problems: Low Oxygen Saturation

    MAUDE report 50744783 ↗

Other Ventilator, emergency, manual (resuscitator) devices

All Ventilator, emergency, manual (resuscitator) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.