YUKON OCT Spinal System 7601-05030M-G1Recalls & reported problems
VB Spine LLC · Polyaxial Screw ML Ø5.0x30 mm · Class 2 · Implant
All models of YUKON OCT Spinal System (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10888857395831), received between Oct 29, 2021 and Nov 17, 2022 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture7
- Migration2
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryNov 17, 2022Fracture; Migration; Fracture; Migration
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY 14 MONTHS AFTER IMPLANTATION TO ADDRESS TWO FRACTURED YUKON POLYAXIAL SCREWS AT T1 AND ONE MIGRATED YUKON SET SCREW. THIS REPORT IS FOR THE SECOND OF TWO FRACTURED YUKON POLYAXIAL SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15821849 ↗InjuryNov 17, 2022Fracture; Migration; Fracture; Migration
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY 8 MONTHS AFTER IMPLANTATION TO ADDRESS TWO FRACTURED YUKON POLYAXIAL SCREWS AT T1, ONE OF WHICH ALSO MIGRATED AND ONE MIGRATED YUKON SET SCREW. THIS REPORT IS FOR THE SECOND OF TWO FRACTURED YUKON POLYAXIAL SCREWS. THIS SCREW ALSO MIGRATED AT T1.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15815864 ↗InjuryNov 17, 2022Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY 3 MONTHS AFTER IMPLANTATION TO ADDRESS TWO FRACTURED YUKON POLYAXIAL SCREWS AT T1. THIS REPORT IS FOR THE SECOND SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15815682 ↗InjurySep 9, 2022Fracture; Fracture
IT WAS REPORTED THAT APPROXIMATELY 3 MONTHS POST-OPERATIVELY A YUKON OCT POLYAXIAL SCREW SHANK FRACTURED AT THE TULIP HEAD AND SHANK JUNCTION AT T1. REVISION SURGERY HAS NOT BEEN PERFORMED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15396559 ↗InjurySep 9, 2022Fracture; Fracture
IT WAS REPORTED THAT APPROXIMATELY 11- 14 MONTHS POST-OPERATIVELY A YUKON OCT POLYAXIAL SCREW SHANK FRACTURED AT THE TULIP HEAD AND SHANK JUNCTION AT T1. REVISION SURGERY HAS NOT BEEN PERFORMED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15396543 ↗InjurySep 9, 2022Fracture; Fracture
IT WAS REPORTED THAT APPROXIMATELY 8 MONTHS POST-OPERATIVELY A YUKON OCT POLYAXIAL SCREW SHANK FRACTURED AT THE TULIP HEAD AND SHANK JUNCTION AT T1. REVISION SURGERY HAS NOT BEEN PERFORMED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15396451 ↗MalfunctionOct 29, 2021Fracture; Fracture
THIS REPORT SUMMARIZES ONE MALFUNCTION EVENT WHERE A YUKON OCT POLYAXIAL SCREW BROKE WHILE BEING REDIRECTED INTO A NEW PATH. THE SCREW SHAFT WAS RETRIEVED AND SURGERY WAS SUCCESSFULLY COMPLETED WITH ANOTHER SCREW AND A 5 MINUTE DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12721080 ↗
Other Orthosis, cervical pedicle screw spinal fixation devices
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- CONNECTOR 79-2002Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- CONNECTOR 79-2003Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- OASYS 48551000Stryker Corporation · Orthosis, cervical pedicle screw spinal fixation45 reports
- SYMPHONY 1020-35-112Medos International Sàrl · Posterior cervical screw system6 deaths15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page