Ventlab AF2142MBRecalls & reported problems
Ventlab, LLC · AirFlow Pediatric Resuscitator with Integrated Manometer, B/V Filter, 40cm Pop-Off, Bag Reservoir, and Mask · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889483090237), received between Apr 3, 2024 and Apr 3, 2024 — classified as 1 death.
The most frequently reported specific device problem is “Connection Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 11, 2023Open, Classified
A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
Z-0361-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Connection Problem1
- Disconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Respiratory Failure1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathApr 3, 2024Disconnection; Connection Problem; Disconnection; Connection Problem
PRODUCT COMPLAINT # (B)(4) REGARDING OUR ITEM # 87-AF2142MBEA- RESUS, PED, AIRFL, MAN, POP-OFF (SUNMED # AF2142MB). OUR CUSTOMER, STEPHANIE EDWARDS, HAS REPORTED "AMBU BAGS / TAPED TO BACK / O2 TUBE THRU INSIDE COMES OUT END HOOK TO O2 SOURCE / TUBE THT GOES TO BACK/EXHAST TEETHIN IS NOT CONNECTED TUBE IS COMING OUT. CUSTOMER STATES THE PAITENT DID EXPIRE BUT DID NOT FEEL BECAUSE OF THIS . THEY HAD TO USE 4 BAGS DUE TO TUBE COMING OUT." PER THE CONVERSATION WITH PRODUCT QUALITY, NO INJURY/DEATH WAS INVOLVED - (B)(6) (COMPLAINTANT).
Patient problems: No Clinical Signs, Symptoms or Conditions; Respiratory Failure
MAUDE report 19038189 ↗
Other Ventilator, Emergency, Manual (Resuscitator) devices
- Ventlab AF5140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall8 reports
- Curaplex AF2000Bound Tree Medical, LLC · Ventilator, emergency, manual (resuscitator)1 recall6 reports
- Ventlab AF5140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall5 reports
- Ventlab AF5142MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab AF2140MBPVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall2 reports
- Ventlab AF1140MBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Ventlab VN3100OBVentlab, LLC · Ventilator, Emergency, Manual (Resuscitator)1 recall1 report
- Medline CPRM1126FPWMMedline Industries, Inc. · Ventilator, emergency, manual (resuscitator)1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.