Skip to content

Ventlab AF2142MBRecalls & reported problems

Ventlab, LLC · AirFlow Pediatric Resuscitator with Integrated Manometer, B/V Filter, 40cm Pop-Off, Bag Reservoir, and Mask · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889483090237), received between Apr 3, 2024 and Apr 3, 2024 — classified as 1 death.

The most frequently reported specific device problem is “Connection Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 11, 2023Open, Classified

    A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

    Z-0361-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Connection Problem1
  • Disconnection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Respiratory Failure1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathApr 3, 2024Disconnection; Connection Problem; Disconnection; Connection Problem

    PRODUCT COMPLAINT # (B)(4) REGARDING OUR ITEM # 87-AF2142MBEA- RESUS, PED, AIRFL, MAN, POP-OFF (SUNMED # AF2142MB). OUR CUSTOMER, STEPHANIE EDWARDS, HAS REPORTED "AMBU BAGS / TAPED TO BACK / O2 TUBE THRU INSIDE COMES OUT END HOOK TO O2 SOURCE / TUBE THT GOES TO BACK/EXHAST TEETHIN IS NOT CONNECTED TUBE IS COMING OUT. CUSTOMER STATES THE PAITENT DID EXPIRE BUT DID NOT FEEL BECAUSE OF THIS . THEY HAD TO USE 4 BAGS DUE TO TUBE COMING OUT." PER THE CONVERSATION WITH PRODUCT QUALITY, NO INJURY/DEATH WAS INVOLVED - (B)(6) (COMPLAINTANT).

    Patient problems: No Clinical Signs, Symptoms or Conditions; Respiratory Failure

    MAUDE report 19038189 ↗

Other Ventilator, Emergency, Manual (Resuscitator) devices

All Ventilator, Emergency, Manual (Resuscitator) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.