Cardica C-Port Flex-A Plus Distal Anastomosis System FG-000150Recalls & reported problems
Aesdex, LLC · C-Port FLEX-A Plus Distal Anastomosis System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 11814900001502), received between Dec 7, 2017 and Apr 13, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Component” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K073123); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Defective Component2
- Failure to Fire2
- Fracture2
- Material Rupture2
- Device Dislodged or Dislocated1
- Retraction Problem1
Patient problems
- Not Applicable2
- Device Embedded In Tissue or Plaque1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 13, 2018Fracture; Material Rupture; Defective Component; Failure to Fire; Fracture; Material Rupture; Defective Component; Failure to Fire
THE C-PORT FLEX-A DEVICE WAS PREPARED WITH A VEIN GRAFT. THE PA ASSISTING IN THE PROCEDURE ACTIVATED THE C-PORT BY TURNING THE KNOB. UPON TURNING THE KNOB THE DEVICE BROKE INTO PIECES THAT SCATTERED AND WAS ACCOMPANIED BY A LOUD NOISE. THERE WAS NO REPORTED ADVERSE EVENT EITHER TO THE PATIENT OR TO USER/SURGICAL PERSONNEL.
Patient problems: Not Applicable
MAUDE report 7431368 ↗MalfunctionApr 9, 2018Fracture; Material Rupture; Defective Component; Failure to Fire; Fracture; Material Rupture; Defective Component; Failure to Fire
DURING DEPLOYMENT OF THE DEVICE AND WHILE BEING HELD BY THE SURGEON, THE DEVICE HANDLE BROKE INTO PIECES, ACCOMPANIED BY A LOUD NOISE. THERE WAS NO REPORTED ADVERSE EVENT EITHER TO THE PATIENT OR TO USER/SURGICAL PERSONNEL.
Patient problems: Not Applicable
MAUDE report 7410270 ↗MalfunctionDec 7, 2017Retraction Problem; Device Dislodged or Dislocated; Retraction Problem; Device Dislodged or Dislocated
THE PATIENT UNDERWENT A ROBOTIC TECAB PROCEDURE IN WHICH THE C-PORT DEVICE WAS USED TO PERFORM THE DISTAL BYPASS ANASTOMOSIS. THE DEVICE FUNCTIONED AS INTENDED AND THE PROCEDURE WAS COMPLETED AS INTENDED. FOLLOWING A POST-OPERATIVE CHEST X-RAY, A SMALL METALLIC DENSITY WAS NOTED IN THE PATIENT'S LEFT CHEST THAT WAS PREVIOUSLY NOT RECOGNIZED. AFTER SURGERY, THE PATIENT WAS TRANSFERRED FOR NORMAL POSTOPERATIVE CARE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THE METALLIC DENSITY.
Patient problems: Device Embedded In Tissue or Plaque; No Known Impact Or Consequence To Patient
MAUDE report 7098161 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K073123. One submission often covers many models, so check the product and lot details in each record.
- Z-2434-2018 · Apr 2, 2018 · TerminatedAesDex, LLC Cardica C-Port Flex-A PLUS Distal Anastomosis System, Catalog Number: FG-000150 The Cardica C-Port Flex-A PLUS Distal Anastomosis System, delivers a series of clips
Other Clip, Implantable devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,808 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,959 reports
- Epix® Universal Clip Applier CA500Applied Medical Resources Corporation · Clip, Implantable1 recall1 death1,183 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,183 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,115 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE791 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death768 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page