MicroCutter 30 White Reload FG-025310Recalls & reported problems
Aesdex, LLC · MicroCutter 30 White Reload is compatible with the MicroCutter 5/80 Stapler · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 11814900003100), received between Apr 5, 2017 and Jun 20, 2017 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Misfire” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Misfire1
Patient problems
- Blood Loss2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 20, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PROCEDURE WAS A ROBOTIC LEFT LOWER LOBECTOMY. THE MICROCUTTER 5/80 STAPLER WAS LOADED WITH A MICROCUTTER 30 WHITE RELOAD AND WAS POSITIONED FOR TRANSECTION OF THE INFERIOR (LOWER LOBE) PULMONARY VEIN. A BRAIDED SILK SUTURE WAS USED TO FACILITATE PLACEMENT OF TISSUE INTO THE DEVICE AND WAS THEN REMOVED PRIOR TO FIRING THE STAPLER. UPON INITIAL INSPECTION, THE STAPLE LINE WAS HEMOSTATIC ON BOTH THE PATIENT AND THE SPECIMEN SIDES OF THE TRANSECTION. THE SURGEON MANIPULATED THE LOBE POST TRANSECTION OF THE INFERIOR PULMONARY VEIN (IPV) AND OOZING WAS NOTED ON THE SPECIMEN SIDE OF THE STAPLE LINE. OOZING WAS CONTROLLED BY USING A COVIDIEN STAPLER. THE PATIENT SIDE OF THE TRANSECTION REMAINED HEMOSTATIC. THE SURGEON CONTINUED WITH THE SURGICAL PROCEDURE INCLUDING TRANSECTING THE SUPERIOR SEGMENTAL ARTERY USING A MICROCUTTER 5/80 STAPLER. THE LUNG SPECIMEN WAS REMOVED FROM THE PATIENT. APPROXIMATELY 45 TO 60 MINUTES AFTER TRANSECTING THE IPV AND AFTER REMOVING THE SPECIMEN, THE LOWER LOBE HILAR REGION BEGAN TO BLEED. THE PATIENT WAS CONVERTED TO AN OPEN THORACOTOMY AND THE SURGEON COMMUNICATED THE AREA OF CONCERN WAS THE INFERIOR PULMONARY VEIN. HE WASN'T ABLE TO VISUALIZE THE STAPLE
Patient problems: Blood Loss
MAUDE report 6655512 ↗MalfunctionApr 5, 2017Misfire; Misfire
THE SURGEON WAS PERFORMING AN OPEN THORACOTOMY. THE MICROCUTTER 5/80 STAPLER AND A MICROCUTTER 30 CURVED TIP BLUE RELOAD WERE USED TO DIVIDE A BASAL BRANCH OF THE PULMONARY ARTERY WITH ACCEPTABLE RESULTS. THE STAPLER AND THE MICROCUTTER 30 WHITE RELOAD WERE THEN USED TO DIVIDE A SMALL SEGMENTAL PULMONARY ARTERY BRANCH WITH RESULTANT BLEEDING NOTED AT THE PATIENT SIDE OF THE STAPLE LINE. THERE DID NOT APPEAR TO BE ANY BLEEDING FROM THE SPECIMEN SIDE OF THE SEGMENTAL PULMONARY ARTERY BRANCH DIVISION. THE SURGEON REPORTED THAT THERE DID NOT SEEM TO BE ANY STAPLES PRESENT IN THE VESSEL. THE SURGEON THEN PLACED SUTURES ON BOTH SIDES OF THE TRANSECTED SEGMENTAL PULMONARY ARTERY BRANCH. A DIFFERENT STAPLER AND DIFFERENT RELOADS WERE USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTED COMPLICATIONS.
Patient problems: Blood Loss
MAUDE report 6462709 ↗
Other Staple, Implantable devices
- Endo GIA EGIA60AMTCovidien LP · STAPLE, IMPLANTABLE19 deaths10,680 reports
- Endo GIA Ultra EGIAUSTNDCovidien LP · STAPLE, IMPLANTABLE5 deaths9,730 reports
- Signia SIGPHANDLECovidien LP · STAPLE, IMPLANTABLE1 recall3 deaths5,356 reports
- ECHELON FLEX PSEE60AEthicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE1 recall13 deaths3,294 reports
- Endo GIA Ultra EGIAUSHORTCovidien LP · STAPLE, IMPLANTABLE2 deaths3,011 reports
- Endo GIA EGIA45AMTCovidien LP · STAPLE, IMPLANTABLE10 deaths2,965 reports
- Endo GIA Ultra EGIAUXLCovidien LP · STAPLE, IMPLANTABLE2,909 reports
- ENDOPATH ATS45Ethicon Endo-Surgery, LLC · STAPLE, IMPLANTABLE2 deaths2,046 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page