AU B96695Recalls & reported problems

Beckman Coulter, Inc. · AU680 Chemistry Analyzer for Direct Track Sampling (with ISE) · Class 1

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 14987666542996), received between Nov 20, 2020 and Dec 3, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (2 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 30, 2025Completed

    Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.

    Z-1908-2025 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Low Test Results2
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUS LOW ION-SELECTIVE ELECTROLYTES (ISE), SPECIFICALLY SODIUM (NA), PATIENT RESULTS ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER STATED SODIUM RESULTS WERE LESS THAN 130 MMOL/L. THE CUSTOMER DID NOT PROVIDE PATIENT DATA PRINTOUTS OR DEMOGRAPHICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20835751 ↗
  • MalfunctionMar 28, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED THE GENERATION OF ERRONEOUS ION-SELECTIVE ELECTROLYTES (ISE) RESULTS FOR EIGHT (8) PATIENT SAMPLES ON THEIR AU680 CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS OR DATA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13931430 ↗
  • MalfunctionNov 20, 2020Low Test Results; Low Test Results

    THE CUSTOMER REPORTED THE GENERATION OF AN ERRONEOUS LOW SODIUM (NA) PATIENT RESULT ON THEIR AU680 INSTRUMENT. THE RESULT WAS NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. THE CUSTOMER VERBALLY PROVIDED EXAMPLES OF THE RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10877884 ↗

Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices

All Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.