AU B96695Recalls & reported problems
Beckman Coulter, Inc. · AU680 Chemistry Analyzer for Direct Track Sampling (with ISE) · Class 1
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 14987666542996), received between Nov 20, 2020 and Dec 3, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 30, 2025Completed
Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.
Z-1908-2025 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Low Test Results2
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 3, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUS LOW ION-SELECTIVE ELECTROLYTES (ISE), SPECIFICALLY SODIUM (NA), PATIENT RESULTS ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER STATED SODIUM RESULTS WERE LESS THAN 130 MMOL/L. THE CUSTOMER DID NOT PROVIDE PATIENT DATA PRINTOUTS OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20835751 ↗MalfunctionMar 28, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED THE GENERATION OF ERRONEOUS ION-SELECTIVE ELECTROLYTES (ISE) RESULTS FOR EIGHT (8) PATIENT SAMPLES ON THEIR AU680 CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS OR DATA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13931430 ↗MalfunctionNov 20, 2020Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF AN ERRONEOUS LOW SODIUM (NA) PATIENT RESULT ON THEIR AU680 INSTRUMENT. THE RESULT WAS NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. THE CUSTOMER VERBALLY PROVIDED EXAMPLES OF THE RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10877884 ↗
Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,197 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall911 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls679 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death694 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death646 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE685 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.