AU B96696Recalls & reported problems
Beckman Coulter, Inc. · AU680 Chemistry Analyzer (with ISE) · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Beckman Coulter AU chemistry analyzer is a laboratory machine that measures patient samples for clinical tests. Of its 129 reports, 127 are malfunctions and 2 are injuries, with no deaths; most describe incorrect test results, many of them falsely low. 1 recall, now closed, started in April 2025 and concerns sample probe tips that can lead to falsely low results.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 129
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 14987666543009), received between Nov 30, 2017 and Jun 3, 2026 — classified as 2 injuries, 127 malfunctions.
The most frequently reported specific device problem is “High Test Results” (55 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 30, 2025Class IICompleted
Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.
Z-1908-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- High Test Results55
- Low Test Results50
- Incorrect, Inadequate or Imprecise Result or Readings23
- High Readings2
- Imprecise or Erratic Results2
- Calibration Problem1
- Low Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions83
- No Consequences Or Impact To Patient42
- No Known Impact Or Consequence To Patient2
- Diaphoresis1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THE AU680 CLINICAL CHEMISTRY GENERATED FALSE LOW PATIENT RESULTS FOR MULTIPLE CHEMISTRIES WHICH INCLUDES SODIUM (NA), POTASSIUM (K),CHLORIDE (CL), CALCIUM (CA), BLOOD UREA NITROGEN (BUN), MAGNESIUM (MG), ETHANOL (ETOH), ASPARTATE AMINOTRANSFERASE (AST), GLUCOSE (GLU), CREATININE (CREA), G-GLUTAMYL TRANSFERASE (GGT), TOTAL PROTEIN (TP), AND CARBON DIOXIDE (CO2) WITH "G" FLAGS. ALSO, THE CUSTOMER REPORTED. THE CUSTOMER ALSO INDICATED THAT THE ISSUE WAS FIRST OBSERVED DURING QUALITY CONTROL (QC) TESTING AND SUBSEQUENTLY AFFECTED PATIENT TESTING, THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE EXACT NUMBER OF ERRONEOUS RESULTS IS UNKNOWN AS PATIENT DATA WAS NOT AVAILABLE. VERBAL EXAMPLE OF RESULTS WAS PROVIDED FOR TWO (2) PATIENTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25374427 ↗MalfunctionMay 18, 2026High Readings
THE CUSTOMER REPORTED OBTAINING ERRONEOUSLY HIGH SODIUM (NA) PATIENT RESULTS AND STABILITY ERROR 3266 - ION SELECTIVE ELECTRODE (ISE) SERUM STANDARD SOLUTION H GENERATED ON THEIR AU680 ISE CLINICAL CHEMISTRY ANALYZER (PRODUCT NUMBER: B96696, SERIAL NUMBER: (B)(6). THERE WERE NO ERRONEOUSLY HIGH NA PATIENT RESULTS RELEASED OUTSIDE THE LABORATORY. THERE WAS NO INJURY, DEATH, OR DELAY/CHANGE IN TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25207228 ↗MalfunctionApr 3, 2026Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW SODIUM PATIENT RESULTS WITH ANION GAP ISSUES ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24778808 ↗MalfunctionApr 2, 2026High Test Results
THE CUSTOMER REPORTED OBTAINING HIGH PATIENT RESULTS WITH MULTIPLE FLAGS ON ION SELECTIVE ELECTRODE (ISE), SODIUM (NA), POTASSIUM (K) AND CHLORIDE (CL) FOR THREE PATIENTS, INVOLVING THE AU680 CLINICAL CHEMISTRY ANALYZER. THE SAMPLES WERE REPEATED ON ANOTHER AU ANALYZER WITH NORMAL RESULTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS BUT PROVIDED RESULTS FOR THREE PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24772220 ↗MalfunctionMar 17, 2026Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW PATIENT AND QUALITY CONTROL RESULTS ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE CUSTOMER DID NOT SPECIFY THE AFFECTED ANALYTES. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24621793 ↗MalfunctionMar 6, 2026Low Test Results
THE CUSTOMER REPORTED THE AU680 CLINICAL CHEMISTRY ANALYZER GENERATED ERRONEOUSLY LOW PATIENT RESULTS FOR SODIUM. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24536635 ↗MalfunctionFeb 10, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW PATIENT RESULTS FOR ALKALINE PHOSPHATASE (ALP), ALANINE AMINOTRANSFERASE (ALT), CARBON DIOXIDE (CO2), CALCIUM ARSENAZO (CALA), TOTAL BILIRUBIN (TBIL) SODIUM (NA), POTASSIUM (K) AND CHLORIDE (CL) ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE SAMPLE WAS REPEATED ON A DIFFERENT AU ANALYZER WITH NORMAL RESULTS. THE LOW RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24308663 ↗MalfunctionFeb 2, 2026Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW OR ZERO PATIENT RESULTS FOR MULTIPLE CHEMISTRIES ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE AFFECTED CHEMISTRY WAS NOT PROVIDED. THE PATIENT SAMPLES WERE REPEATED ON A DIFFERENT AU ANALYZER WITH NORMAL RESULTS. THE CUSTOMER DID NOT RELEASE THE RESULTS OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24242290 ↗MalfunctionJan 15, 2026Low Readings
THE CUSTOMER REPORTED LOW PATIENT RESULTS ON MULTIPLE ANALYTES WITH G FLAGS INCLUDING AST (ASPARTATE AMINOTRANSFERASE), ALT (ALANINE AMINOTRANSFERASE), GLU (GLUCOSE), NA (SODIUM) K (POTASSIUM) AND CL (CHLORIDE) ON THEIR AU680 CHEMISTRY ANALYZER PART NUMBER B96696 AND INSTRUMENT SERIAL NUMBER (B)(6). THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, DEATH, OR DELAY/CHANGE IN TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24083344 ↗MalfunctionDec 16, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW ION SELECTIVE ELECTRODE (ISE) FOR SODIUM (NA), POTASSIUM (K) AND CHLORIDE (CL) ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE PATIENT RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE SAMPLES WERE RE-ANALYZED WITH NORMAL RESULTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. PROVIDED THE RESULTS FOR TWO PATIENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23819502 ↗MalfunctionDec 4, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW SODIUM (NA) PATIENT RESULTS ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23720682 ↗MalfunctionNov 24, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY HIGH ION-SELECTIVE ELECTROLYTES (ISE) PATIENT RESULTS, INCLUDING SODIUM (NA), CHLORIDE (CL) AND POTASSIUM (K), ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. THE CUSTOMER PROVIDED DATA FOR ONE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23638172 ↗MalfunctionNov 12, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED RECOVERING LOW GLUCOSE (GLU) AND ION SELECTIVE ELECTRODE (ISE) RESULTS, WHICH INCLUDED SODIUM (NA) WHEN RUN ON THEIR AU680 CHEMISTRY ANALYZER (PN B96696, SN (B)(6)). THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23533960 ↗MalfunctionOct 21, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE AU680 CLINICAL CHEMISTRY ANALYZER GENERATED APPROXIMATELY THIRTEEN (13) ERRONEOUSLY LOW PATIENT SAMPLE RESULTS FOR SODIUM (NA). THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23344563 ↗MalfunctionAug 12, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED OBTAINING THREE PATIENT SAMPLE RESULTS ERRONEOUSLY HIGH FOR SODIUM (NA) ON THE ION SELECTIVE ELECTRODE (ISE) MODULE OF THE AU680 ANALYZER (PN B96696, SN (B)(6)). ONE OF THESE AFFECTED PATIENT SAMPLE RESULT WAS RELEASED FROM THE LABORATORY. THE CUSTOMER SUBSEQUENTLY RE-TESTED THE SAMPLE ON A DIFFERENT ANALYZER, OBTAINING ACCEPTABLE RESULTS, AND SUBMITTED A CORRECTIVE ACTION REPORT. THERE WAS NO REPORT OF CHANGES TO TREATMENT, INJURY OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDED PATIENT DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22790057 ↗MalfunctionJul 11, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW PATIENT RESULTS FOR TOTAL BILIRUBIN (TBIL), DIRECT BILIRUBIN (DBIL), C-REACTIVE PROTEIN (CRP), AND ALANINE AMINOTRANSFERASE (ALT), WITH A ¿G¿ FLAG, ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE ANALYZER ALSO DISPLAYED MULTIPLE ERRORS, INCLUDING PHOTOCAL DATA ERRORS, CUVETTE MEAN CHECK ERRORS, AND PHOTOMETRY ERRORS. ADDITIONALLY, THE MID-STANDARD AND BUFFER SOLUTIONS WERE INDICATED AS EMPTY WITHOUT GENERATING THE EXPECTED WARNINGS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22481469 ↗MalfunctionJun 12, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED SIX PATIENTS HAD HIGH SODIUM RESULTS WITH A ¿PH¿ FLAG ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE SAMPLES WERE REPEATED ON ANOTHER AU ANALYZER, RECOVERING NORMAL RESULTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. THE CUSTOMER VERBALLY PROVIDED RESULTS FOR ONLY ONE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22197670 ↗MalfunctionMay 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THE GENERATION OF ERRATIC GLUCOSE PATIENT RESULTS ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21970381 ↗MalfunctionMar 28, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED THE AU680 CLINICAL CHEMISTRY ANALYZER GENERATED ERRONEOUSLY LOW PATIENT RESULTS WITH 'G' FLAGS ON BLOOD UREA NITROGEN (BUN), CREATININE (CRE), CHLORIDE (CL), BICARBONATE (CO2), CALCIUM (CA), TOTAL PROTEIN (TP), ALBUMIN (ALB), ASPARTATE AMINOTRANSFERASE (AST), ALANINE AMINOTRANSFERASE (ALT), ALKALINE PHOSPHATASE (ALP) AND TOTAL BILIRUBIN (TBIL) FOR MULTIPLE PATIENTS. THE CUSTOMER PROVIDED EXAMPLE OF FALSE RESULTS FOR TWO (2) PATIENTS. THE CUSTOMER DID NOT PROVIDE PATIENT DEMOGRAPHICS. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21721100 ↗MalfunctionMar 25, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER QUESTIONED ANION GAP (AGAP) ON PATIENT RESULTS GENERATED ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21694257 ↗MalfunctionFeb 18, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING FALSE LOW ISE (ION SELECTIVE ELECTRODE) PATIENT RESULTS WHICH INCLUDE SODIUM (NA), CHLORIDE (CL), AND POTASSIUM (K) ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS, AND THE NUMBER OF AFFECTED PATIENT SAMPLES IS UNKNOWN. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21403634 ↗MalfunctionFeb 18, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING TWO (2) FALSE LOW ISE (ION SELECTIVE ELECTRODE) PATIENT RESULTS WHICH INCLUDE SODIUM (NA), CHLORIDE (CL), AND POTASSIUM (K) ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS, AND THE NUMBER OF AFFECTED PATIENT SAMPLES IS UNKNOWN. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21400393 ↗MalfunctionFeb 5, 2025Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING FALSE LOW ISE (ION SELECTIVE ELECTRODE) PATIENT RESULTS WHICH INCLUDE SODIUM (NA), CHLORIDE (CL), AND POTASSIUM (K) ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS, AND THE NUMBER OF AFFECTED PATIENT SAMPLES IS UNKNOWN. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21318218 ↗MalfunctionJan 10, 2025Low Test Results; Low Test Results
CUSTOMER REPORTED THE AU680 CHEMISTRY ANALYZER GENERATED ERRONEOUS LOW ION SELECTIVE ELECTRODE (ISE), SODIUM (NA), CHLORIDE (CL), POTASSIUM (K), ALBUMIN (ALB), TOTAL PROTEIN (TP), AND ALKALINE PHOSPHATASE (ALP) PATIENT RESULTS. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED TO THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21132254 ↗MalfunctionNov 18, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW QUALITY CONTROL AND PATIENT RESULTS FOR ION SELECTIVE ELECTRODE ON SODIUM, POTASSIUM AND CHLORIDE ON THEIR AU680 CLINICAL CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20712399 ↗
FDA clearance and classification
The FDA regulates product code JJE (analyzer, chemistry (photometric, discrete), for clinical use) under 21 CFR 862.2160 as a Class I device. 21 CFR 862.2160 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the AU B96696
Has the AU B96696 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the company has completed it but the FDA has not yet closed it. The recall, Z-1908-2025 (Class II), began on Apr 30, 2025: Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the AU B96696?
The FDA's MAUDE database lists 129 adverse event reports for this device, received between Nov 30, 2017 and Jun 3, 2026. In those reports, the most common specific device problems are “High Test Results” (55), “Low Test Results” (50) and “Incorrect, Inadequate or Imprecise Result or Readings” (23); the most common patient problem is “Diaphoresis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the AU B96696?
None of the reports are classified as deaths, 2 as injuries and 127 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the AU B96696 FDA approved?
Devices of this type (product code JJE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Beckman Coulter, Inc.
Who makes the AU B96696, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand AU, model B96696). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Analyzer, Chemistry (Photometric, Discrete), For Clinical Use devices
- Alinity 03R6701Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death1,184 reports
- Alinity 03R6501Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall903 reports
- Atellica CH 930 Analyzer 11067000Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE5 recalls677 reports
- DxI 9000 Access Immunoassay Analyzer C11137Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use17 recalls1 death3 reports
- ARCHITECT 3L77-01Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE2 recalls1 death692 reports
- ARCHITECT 3M74-02Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall650 reports
- ARCHITECT 2P24-40Abbott Laboratories · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall1 death645 reports
- cobas e 411 Immunoassay Analyzer 04775201001Roche Diagnostics GmbH · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE680 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page