MicroScan B1018-284Recalls & reported problems
Beckman Coulter, Inc. · WalkAway-96 plus Instrument · Class 2
All models of MicroScan (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15099590658687), received between Aug 8, 2019 and Nov 6, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Handling Problem” (1 report).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Oct 21, 2019Class IIOpen, Classified
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
Z-1084-2022 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- Sep 4, 2019Class IITerminated
Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
Z-1371-2021 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Handling Problem1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- Injury1
- Septic Shock1
- Urinary Tract Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 6, 2019Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED ISSUES WITH LOW PROBABILITY IDENTIFICATIONS AND (B)(6) RESISTANT MISREADS WHEN COMPARED TO VISUAL READ. CUSTOMER INDICATED THEY HAD TWO SPECIMENS WHERE ERRONEOUS RESULTS WERE REPORTED. FIRST SPECIMEN REPORTED AS (B)(6). UPON REPEAT, THE ISOLATE IDENTIFIED AS (B)(6). THE (B)(6) RESULT WAS REPORTED OUT AND QUESTIONED. THE PATIENT HAD A SPECIMEN RUN SEVERAL DAYS PRIOR THAT RESULTED AS (B)(6). THE (B)(6) RESULT WAS OBTAINED ON (B)(6) 2019. PHYSICIAN BEGAN (B)(6) TREATMENT FOR THE (B)(6) YEARS OLD FEMALE DIAGNOSED WITH SEPTIC SHOCK. AFTER THE REPEAT RESULT OF (B)(6) WAS OBTAINED ON (B)(6) 2019, THE CUSTOMER INDICATED THE TREATMENT WAS GOING TO BE CHANGED TO (B)(6). SECOND SPECIMEN WAS VERY RESISTANT AND TESTED AS ESCHERICHIA COLI ESBL NEGATIVE ON (B)(6) 2019. UPON REPEAT ON (B)(6) 2019, THE ORGANISM TESTED AS ESBL NEGATIVE AGAIN BUT WAS MANUALLY EDITED TO POSITIVE BASED ON VISUAL READ OF THE PANEL. THE PATIENT WAS A (B)(6) YEARS OLD MALE WITH A URINARY TRACT INFECTION. CUSTOMER DID NOT KNOW WHAT THE INITIAL TREATMENT WAS; THE AMENDED TREATMENT WAS MEROPENEM. THERE WAS A THIRD SPECIMEN HOWEVER NO ERRONEOUS RESULTS WERE REPORTED. THE ISOLATE TESTED AS KLEBSIELLA OXYTOCA
Patient problems: Septic Shock; Urinary Tract Infection
MAUDE report 9287658 ↗InjuryAug 8, 2019Device Handling Problem; Device Handling Problem
WHILE ONSITE ON (B)(6) 2018 CONDUCTING A REPAIR ON WALKAWAY (WA) 96 PLUS, SERIAL NUMBER (B)(4), THE BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) INJURED HIS SHOULDER.
Patient problems: Injury
MAUDE report 8876535 ↗
FDA clearance and classification
The FDA regulates product code LRG (instrument for auto reader & interpretation of overnight suscept. systems) under 21 CFR 866.1640 as a Class II device. 21 CFR 866.1640 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the MicroScan B1018-284
Has the MicroScan B1018-284 been recalled?
Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-1084-2022 (Class II), began on Oct 21, 2019: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the MicroScan B1018-284?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Aug 8, 2019 and Nov 6, 2019. In those reports, the most common specific device problems are “Device Handling Problem” (1) and “Incorrect, Inadequate or Imprecise Result or Readings” (1); the most common patient problems are “Injury” (1), “Septic Shock” (1) and “Urinary Tract Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MicroScan B1018-284?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the MicroScan B1018-284, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand MicroScan, model B1018-284). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems devices
- MicroScan B1018-440Beckman Coulter, Inc. · Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems1 recall1 death2 reports
- MicroScan B1018-496Beckman Coulter, Inc. · Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems1 recall1 report
- Thermo Scientific™ Sensititre™ ARIS HiQ™ System V4000Trek Diagnostic Systems Limited · Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page