MicroScan B1018-496Recalls & reported problems
Beckman Coulter, Inc. · DxM 1096 MicroScan Walkaway Instrument · Class 2
All models of MicroScan (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15099590690892), received between Oct 3, 2024 and Oct 3, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Oct 21, 2019Class IIOpen, Classified
WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
Z-1086-2022 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
Most reported problems
Device problems
- Therapeutic or Diagnostic Output Failure1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 3, 2024Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
THE CUSTOMER REPORTED THE GENERATION OF HIGH PROBABILITY MISIDENTIFICATION RESULTS USING B1017-219 PC45 PANELS PROCESSED ON THEIR WALKAWAY (WA) SYSTEM (S/N: (B)(6). THE CUSTOMER STATED THAT A PATIENT ISOLATE WAS INITIALLY IDENTIFIED AS S. SHOLIFERI (BIOTYPE 313731, 99.99% PROBABILITY) USING PC45 PANEL, AND THIS IDENTIFICATION WAS REPORTED TO THE PHYSICIAN. REPEAT TESTING USING BIOFIRE AND B1017-166 RPID2 PANEL WAS PERFORMED, AND BOTH GAVE AN IDENTIFICATION OF S. AUREUS. THE UPDATED IDENTIFICATION REPORT WAS PROVIDED TO THE PHYSICIAN, AND THE PHYSICIAN HAD TO CHANGE THE PATIENT'S COURSE OF TREATMENT BASED ON THE AMENDED REPORT WITH THE CORRECT ORGANISM IDENTIFICATION, HOWEVER NO ADDITIONAL DETAILS WERE PROVIDED. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.
Patient problems: Insufficient Information
MAUDE report 20364584 ↗
FDA clearance and classification
The FDA regulates product code LRG (instrument for auto reader & interpretation of overnight suscept. systems) under 21 CFR 866.1640 as a Class II device. 21 CFR 866.1640 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the MicroScan B1018-496
Has the MicroScan B1018-496 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1086-2022 (Class II), began on Oct 21, 2019: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the MicroScan B1018-496?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Oct 3, 2024. That report lists device problem “Therapeutic or Diagnostic Output Failure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MicroScan B1018-496?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the MicroScan B1018-496, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand MicroScan, model B1018-496). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems devices
- MicroScan B1018-284Beckman Coulter, Inc. · Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems2 recalls2 reports
- MicroScan B1018-440Beckman Coulter, Inc. · Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems1 recall1 death2 reports
- Thermo Scientific™ Sensititre™ ARIS HiQ™ System V4000Trek Diagnostic Systems Limited · Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page