Change Healthcare Cardiology Hemo 14.3.2Recalls & reported problems
Change Healthcare Canada Company · Class 2
All models of Change Healthcare Cardiology Hemo (6)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 3
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
3 FDA recall records name this device's identifier, 3 still open.
7 other recalls cover devices cleared under the same FDA submission (K131497); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 22, 2025Class IIOpen, Classified
Due to complaints, software update may cause software to unexpectedly shutdown.
Z-2071-2025 · Root cause: Software design · Recall details · FDA record ↗
- Dec 2, 2024Class IIOpen, Classified
Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s).
Z-0958-2025 · Root cause: Software design · Recall details · FDA record ↗
- Aug 21, 2024Class IIOpen, Classified
A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
Z-0126-2025 · Root cause: Use error · Recall details · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131497. One submission often covers many models, so check the product and lot details in each record.
- Z-0957-2025 · Dec 2, 2024 · Open, ClassifiedMcKesson Cardiology Hemo software
- Z-0128-2025 · Aug 21, 2024 · Open, ClassifiedMcKesson Cardiology Hemo. Versions 13.0, 13.1.1, 13.1.2, 13.2, 13.2.1, 14.0.
- Z-1489-2019 · Mar 25, 2019 · TerminatedHorizon Cardiology Hemo
- Z-1490-2019 · Mar 25, 2019 · TerminatedMcKesson Cardiology Hemo
- Z-2968-2018 · Mar 12, 2018 · TerminatedMcKesson Cardiology Hemo, Release SW version 13.0. Used for physiological monitoring, image and data processing. McKesson Cardiology Hemo is intended for complete physiological/h
- Z-0042-2015 · Aug 27, 2014 · TerminatedMcKesson Cardiology" Hemo intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.
- Z-0093-2015 · Jun 25, 2014 · TerminatedMcKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, an
Other Computer, Diagnostic, Programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- CARESCAPE CENTRAL STATION MAI800Ge Medical Systems Information Technologies, Inc. · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page