Skip to content

CONMED SCOPE SAVER 100429Recalls & reported problems

Conmed Corporation · CONMED SCOPE SAVER BITE BLOCK, 14 X 20 MM · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20653405050275), received between Feb 9, 2026 and Feb 10, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 10, 2026Material Fragmentation

    A CUSTOMER REPORTED THAT THE 100429, CONMED SCOPE SAVER BITE BLOCK, 14 X 20 MM DEVICE WAS USED IN AN UNKNOWN PROCEDURE ON (B)(6) 2026, AND ¿IT WAS OBSERVED THAT THE BITE BLOCK'S PERFORATED STRAP HAD A RESIDUAL FLAP OF MATERIAL THAT BECAME DISLODGED AND WAS FOUND INSIDE THE PATIENT. THIS INCIDENT OCCURRED TWICE IN ONE DAY.¿. THE PROCEDURE WAS COMPLETED AND THERE WAS NO REPORTED IMPACT OR INJURY TO THE PATIENT. FURTHER ASSESSMENT FOUND "THAT THE INCIDENT OCCURRED WITH 2 SEPARATE PATIENTS. NO INSTRUMENTS WERE NEEDED TO RETRIEVE THE PIECE OF MATERIAL, AND NO BAD OUTCOMES OCCURRED". THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24302714 ↗
  • MalfunctionFeb 9, 2026Material Fragmentation

    A CUSTOMER REPORTED THAT THE 100429, CONMED SCOPE SAVER BITE BLOCK, 14 X 20 MM DEVICE WAS USED IN AN UNKNOWN PROCEDURE ON (B)(6) 2026, AND ¿IT WAS OBSERVED THAT THE BITE BLOCK'S PERFORATED STRAP HAD A RESIDUAL FLAP OF MATERIAL THAT BECAME DISLODGED AND WAS FOUND INSIDE THE PATIENT. THIS INCIDENT OCCURRED TWICE IN ONE DAY.¿. THE PROCEDURE WAS COMPLETED AND THERE WAS NO REPORTED IMPACT OR INJURY TO THE PATIENT. FURTHER ASSESSMENT FOUND "THAT THE INCIDENT OCCURRED WITH 2 SEPARATE PATIENTS. NO INSTRUMENTS WERE NEEDED TO RETRIEVE THE PIECE OF MATERIAL, AND NO BAD OUTCOMES OCCURRED". THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24291687 ↗

Other BRONCHOSCOPE ACCESSORY devices

All BRONCHOSCOPE ACCESSORY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.