INSPIRA AIR BC0524AZRecalls & reported problems
Acclarent, Inc. · INSPIRA AIR BALLOON DILATION SYSTEM 5x24, 1 PK · Class 2
All models of INSPIRA AIR (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705031063089), received between Jul 3, 2023 and Jul 3, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K090660); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
- Use of Device Problem1
Patient problems
- Hypoxia1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 3, 2023Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
THE EVENT WAS REPORTED VIA VOLUNTARY MEDWATCH REPORT (MW5118360). A PEDIATRIC PATIENT UNDERWENT DIRECT LARYNGOSCOPY, BRONCHOSCOPY, AND BALLOON DILATION UNDER INTERMITTENT APNEA FOR SUBGLOTTIC STENOSIS. A 5MM X 24MM INSPIRA AIR BALLOON DILATION SYSTEM (BC0524AZ / LOT# UNKNOWN) WAS USED. THE DEVICE INITIALLY FAILED TO INFLATE THE AIRWAY AND THE PATIENT¿S OXYGEN SATURATION BEGAN TO DROP. THE INFLATION DEVICE HAD BEEN ATTACHED TO THE STYLET PORT AND NOT THE BALLOON PORT. THE SOLUTION HAD BEEN IRRIGATED INTO THE PATIENT¿S AIRWAY. THE PATIENT¿S AIRWAY WAS SUCTIONED AND THE PATIENT REQUIRED BAG AND MASK VENTILATION. THE COMPLAINT DEVICE IS NOT AVAILABLE FOR RETURN. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Hypoxia
MAUDE report 17253213 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090660. One submission often covers many models, so check the product and lot details in each record.
- Z-1861-2013 · Jan 11, 2013 · TerminatedInspira AIR Balloon Dilation system. Size 7 x 24 mm, Syringe volume 6-8 cc. Dilation of airway tree.
- Z-1866-2013 · Jan 11, 2013 · TerminatedInspira AIR Balloon Dilation system. Size 16 x 40 mm, Syringe volume 12 cc. Dilation of airway tree.
- Z-1860-2013 · Jan 11, 2013 · TerminatedInspira AIR Balloon Dilation system. Size 5 x 24 mm, Syringe volume 6-8 cc. Dilation of airway tree.
- Z-1864-2013 · Jan 11, 2013 · TerminatedInspira AIR Balloon Dilation system. Size 12 x 40 mm, Syringe volume 12 cc. Dilation of airway tree.
- Z-1862-2013 · Jan 11, 2013 · TerminatedInspira AIR Balloon Dilation system. Size 8.5 x 24 mm, Syringe volume 12 cc. Dilation of airway tree.
- Z-1865-2013 · Jan 11, 2013 · TerminatedInspira AIR Balloon Dilation system. Size 14 x 40 mm, Syringe volume 12 cc. Dilation of airway tree.
- Z-1863-2013 · Jan 11, 2013 · TerminatedInspira AIR Balloon Dilation system. Size 10 x 40 mm, Syringe volume 12 cc. Dilation of airway tree.
Other BRONCHOSCOPE ACCESSORY devices
- ViziShot 2 FLEX NA-U403SX-4019Gyrus ACMI, LLC · Bronchoscope accessory1 recall349 reports
- PeriView FLEX NA-403D-2021Gyrus ACMI, LLC · Bronchoscope accessory54 reports
- INSPIRA AIR BC1040AZAcclarent, Inc. · BRONCHOSCOPE ACCESSORY1 recall0 reports
- CRE Pulmonary M00550340Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY26 reports
- CRE Pulmonary M00550310Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY21 reports
- CRE Pulmonary M00550350Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY21 reports
- CRE Pulmonary M00550330Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY20 reports
- CRE Pulmonary M00550320Boston Scientific Corporation · BRONCHOSCOPE ACCESSORY15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page