Skip to content

MAXI LD 4171460SRecalls & reported problems

Cordis US Corp. · MAXI LD 7F 14X6 80CM · Class 2

All models of MAXI LD (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032002452), received between Apr 3, 2023 and Jun 9, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Detachment of Device or Device Component1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2025Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    AS REPORTED, THE BALLOON OF A 14MM X 6CM MAXI LD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WAS CAUGHT DURING THE PROCEDURE AND FELL OFF. THE BALLOON WAS CAUGHT DURING REMOVAL FROM A NON-CORDIS STENT AND THE SEPARATION OCCURRED WHILE TRYING TO WITHDRAW THE BALLOON CATHETER. THERE WERE ATTEMPTS TO RETRIEVE THE SEPARATED SEGMENT FROM THE PATIENT USING AN UNKNOWN DEVICE; HOWEVER, THE RETRIEVAL WAS UNSUCCESSFUL, AND THE PATIENT WAS TRANSPORTED BY AMBULANCE TO THE HOSPITAL FOR SURGERY. THE SEPARATED SEGMENT WAS SUCCESSFULLY REMOVED DURING THE SURGICAL INTERVENTION. THE TARGET VESSEL WAS THE ILIAC VEIN, WHICH WAS NEITHER CALCIFIED NOR TORTUOUS. THE SEPARATION OCCURRED IN THE ILIAC VEIN. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS OR THE PROTECTIVE BALLOON COVER. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22168095 ↗
  • MalfunctionApr 3, 2023Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, DURING THE USE OF A 14X6 80CM MAXI LD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER, THE BALLOON IMMEDIATELY RUPTURED DURING AN ATTEMPTED INFLATION AND THE BALLOON WAS UNABLE TO BE PRESSURIZED. AS A RESULT, A NEW 14X6 80CM MAXI LD PTA BALLOON CATHETER WAS USED AS A REPLACEMENT. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE DEVICE WAS BEING USED TO POST DILATE AN UNKNOWN VENOUS STENT. THERE WAS NO DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. A 1:2 CONTRAST TO SALINE RATIO WAS USED TO PREP THE BALLOON. THE DEVICE WAS ABLE TO BE REMOVED EASILY FROM THE PATIENT AND REMAINED IN ONE PIECE DURING ITS REMOVAL. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16665737 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page