MAXI LD 4171840SRecalls & reported problems
Cordis US Corp. · MAXI LD 7F 18X4 80CM · Class 2
All models of MAXI LD (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20705032002605), received between Jul 31, 2017 and Aug 16, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel3
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 16, 2022Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON OF AN 18MM X 40MM 80CM MAXI LD PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED DURING AND INFLATION TO 3 ATMOSPHERES (ATM). AS A RESULT, THE 8MM X 40MM 80CM MAXI LD PTA BALLOON CATHETER WAS REPLACED WITH 16MM X 4MM MAXI LD BALLOON CATHETER WHICH WAS USED WITHOUT ISSUE TO CONTINUE THE PROCEDURE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A LESION IN THE ILIAC VEIN. THERE WAS NO CALCIFICATION OR TORTUOSITY AT THE TARGET LESION, AND THE LESION WAS NOT A CHRONIC TOTAL OCCLUSION (CTO). THE DEVICE WAS STORED AND USED PER THE INSTRUCTIONS FOR USE (IFU) AND MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. A 50/50 CONTRAST AND SALINE MIXTURE WAS USED TO PREP THE BALLOON. THERE WAS NO DIFFICULTY REMOVING THE BALLOON HOOP OR PROTECTIVE COVER DURING PREPARATION. THERE WAS NO DIFFICULTY EXPERIENCED DURING INSERTION OF THE DEVICE, OR WHILE CROSSING THE LESION. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND DURING THIS PROCEDURE, AND THE DEVICE DID NOT KINK DURING ITS USE. THE DEVICE WAS ABLE TO BE REMOVED EASILY FROM THE PATIENT AND REMAINED IN ONE PIECE DURING REMOVAL. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15232632 ↗MalfunctionJul 31, 2017Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE MAXI LD BALLOONS CATHETERS BURST AT TWO ATMOSPHERES (2 ATMS). THERE WERE NO PATIENT INJURY. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6755630 ↗MalfunctionJul 31, 2017Burst Container or Vessel; Burst Container or Vessel
IT WAS REPORTED THAT THE MAXI LD BALLOONS CATHETERS BURST AT TWO ATMOSPHERES (2 ATMS). THERE WERE NO PATIENT INJURY. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6755576 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page