AVIATOR 4244015WRecalls & reported problems

Cordis US Corp. · AVIATOR PLUS .014 4.0X15 142CM · Class 2

All models of AVIATOR (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20705032005736), received between Mar 30, 2018 and Jun 7, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel2
  • Material Rupture1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 7, 2024Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, THE .014 4.0X15 AVIATOR PLUS WAS USED. HOWEVER, IT RUPTURED AT ITS NOMINAL PRESSURE. THEREFORE, IT WAS REPLACED WITH A NEW NON-CORDIS BALLOON CATHETER AND THE PROCEDURE WAS COMPLETED. THE PRODUCT WAS REMOVED INTACT FROM THE PATIENT. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE DEVICE WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, STYLET, OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS, OR THE DAMAGES NOTED PRIOR TO INSERTING THE DEVICE INTO THE PATIENT. THE DEVICE WAS PREPPED PER THE IFU. THE DEVICE WAS PREPPED NORMALLY AND WAS ABLE TO MAINTAIN NEGATIVE PRESSURE. THE LESION WAS IN THE RENAL ARTERY WHICH HAD A 80% STENOSIS. THE LESION WAS NOTED TO HAVE MODERATE CALCIFICATION WITH MILD VESSEL TORTUOSITY. THE DEVICE WAS NOT BEING USED TO TREAT A CHRONIC TOTAL OCCLUSION (CTO). THERE WAS NO RESISTANCE/ FRICTION WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE. THERE WAS NO RESISTANCE/ FRICTION WHILE INSERTING THE BALLOON THROUGH THE UNKNOWN GUIDE CATHETER. THERE WAS NO DIFFICULTY ADVANCING THE BALLOON CATHETER THROUGH THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19491918 ↗
  • MalfunctionNov 20, 2019Material Rupture; Material Rupture

    AS REPORTED, THE .014 4.0 X 15 142CM AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER (BC) RUPTURED AT TWELVE ATMOSPHERES (12ATM) DURING ITS INITIAL INFLATION. THE PROCEDURE WAS THEREFORE COMPLETED USING ANOTHER BALLOON CATHETER AND A CUTTING CATHETER. THERE WAS NO REPORTED PATIENT INJURY. THE LESION WAS AT THE RIGHT POPLITEAL ARTERY WHICH NOTED TO HAVE SEVERE CALCIFICATION AND EXTREMELY SHORT SEVERE STENOSIS. A CONTRALATERAL APPROACH WAS MADE USING A NON-CORDIS GUIDING CATHETER. A NON-CORDIS GUIDEWIRE HAD CROSSED THE LESION. THE DEVICE WILL NOT BE RETURNED AS IT WAS DISCARDED BY MISTAKE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9349777 ↗
  • MalfunctionMar 30, 2018Burst Container or Vessel; Burst Container or Vessel

    DURING USE, IT WAS REPORTED THAT THE AVIATOR PLUS BALLOON CATHETER RUPTURED AT TEN ATMOSPHERE (10 ATM). THERE WAS NO PATIENT INJURY. THE AVIATOR BALLOON WAS REPLACED WITH A NEW UNKNOWN CUTTING BALLOON CATHETER AND THE PROCEDURE WAS COMPLETE SUCCESSFULLY. THE LESION WAS THE LEFT POPLITEAL ARTERY. THE LESION HAD HEAVY STENOSIS AND SEEMED TO BE HEAVILY CALCIFIED. A 6F NON CORDIS SHEATH WAS INSERTED TO MAKE AN IPSILATERAL APPROACH AND A NON CORDIS GUIDEWIRE CROSSED THE LESION. THEN AN AVIATOR PLUS WAS DELIVERED TO THE LESION AND INFLATED. THE LESION WAS THE LEFT POPLITEAL ARTERY. THE LESION HAD HEAVY STENOSIS AND SEEMED TO BE HEAVILY CALCIFIED. THE DEVICE HAS BEEN ALREADY DISCARDED IN THE HOSPITAL.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7384785 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →