Skip to content

AVIATOR 4245040WRecalls & reported problems

Cordis US Corp. · AVIATOR PLUS .014 5.0X40 142CM · Class 2

All models of AVIATOR (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20705032005842), received between Jul 15, 2019 and Apr 10, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel4
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, THE BALLON OF A 5MM X 40MM AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT 5 ATMOSPHERES (ATM). A 4MM X 40MM AVIATOR PLUS PTA BALLOON CATHETER WAS USED AS A REPLACEMENT AND WAS INFLATED TO 10ATM WITHOUT ISSUE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING AN ENDOVASCULAR THERAPY (EVT) PROCEDURE TO TREAT A SUPERFICIAL FEMORAL ARTERY (SFA) LESION. A CONTRALATERAL APPROACH WAS MADE WITH A 6F NON-CORDIS SHEATH. AN UNKNOWN GUIDEWIRE WAS USED TO CROSS THE LESION PRIOR TO USE OF THE AVIATOR PLUS PTA BALLOON CATHETER. A NON-CORDIS BALLOON CATHETER WAS ALSO USED DURING THIS PROCEDURE. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21816179 ↗
  • MalfunctionDec 16, 2021Burst Container or Vessel; Burst Container or Vessel

    A .014 5.0X40 142CM AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT NOMINAL PRESSURE DURING USE. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13017960 ↗
  • MalfunctionApr 14, 2021Leak/Splash; Leak/Splash

    AS REPORTED, THE DOCTOR PROCEEDED TO INFLATE THE BALLOON BUT THE .014 5.0 X 40 142CM AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WOULD NOT HOLD PRESSURE AND LEAKAGE WAS NOTED FROM THE BALLOON. THEREFORE, THE BALLOON WAS REMOVED. THERE WAS NO HARM AND NO REPORTED PATIENT INJURY. THE DEVICE WAS USED IN PATIENT. THE DEVICE WAS STORED AS PER LABELING AND OPENED IN STERILE FIELD. THE BALLOON WAS PREPPED AS PER THE INSTRUCTIONS FOR USE (IFU). THE DEVICE WILL BE RETURNED FOR ANALYSIS. THE EVENT OCCURRED INSIDE THE PATIENT DURING A PROCEDURE IN THE CAROTID ARTERY. THERE WAS NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. NO KINKS OR OTHER DAMAGES WERE NOTED PRIOR TO INSERTING THE PRODUCT THE PRODUCT INTO THE PATIENT. THE DEVICE PREPPED NORMALLY. THE CONTRAST TO SALINE RATIO WAS 50/50. AN INDEFLATOR WAS USED AS THE INFLATION DEVICE AND THE SAME INDEFLATOR WAS USED SUCCESSFULLY WITH OTHER DEVICES. THE PLUNGER DEPRESSED INTO THE SYRINGE/INDEFLATOR WHEN TRYING TO INFLATE. ADDITIONAL PATIENT AND PROCEDURAL DETAILS WERE REQUESTED BUT WERE NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11663935 ↗
  • MalfunctionMay 11, 2020Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, A RUPTURE WAS FOUND ON THE BALLOON OF A 5.0MM X 40MM 142CM .014¿ AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER DURING PREP. THEREFORE, THE PROCEDURE WAS COMPLETED WITH ANOTHER AVIATOR PLUS PTA BALLOON. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED AND HANDLED AS PER THE INSTRUCTION FOR USE (IFU). THERE WERE NO DAMAGES NOTED TO THE PACKAGING OF THE DEVICE PRIOR TO USE. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP OR THE PROTECTIVE BALLOON COVER. THERE WAS NO DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT THE PRODUCT INTO THE PATIENT. THE DEVICE WAS PREPPED NORMALLY (I.E. MAINTAIN NEGATIVE PRESSURE). THE TARGET LESION WAS THE RENAL ARTERY. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10043416 ↗
  • MalfunctionJul 15, 2019Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, THIS WAS AN ENDOVASCULAR TREATMENT (EVT) CASE, AN AVIATOR PLUS 014 5.0X40 142CM WAS USED FOR A PRE-DILATATION, HOWEVER IT RUPTURED AT ITS NOMINAL PRESSURE. THERE WERE NO REPORTED PATIENT INJURIES. THEREFORE IT WAS REPLACED WITH A NEW BALLOON CATHETER AND THE PROCEDURE WAS COMPLETED. ADDITIONAL PROCEDURAL DETAILS WERE REQUESTED BUT ARE UNKNOWN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8790688 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page