AVIATOR 4246015WRecalls & reported problems
Cordis US Corp. · AVIATOR PLUS .014 6.0X15 142CM · Class 2
All models of AVIATOR (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20705032005897), received between Nov 16, 2020 and Feb 18, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Difficult to Remove1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 18, 2021Material Rupture; Material Rupture
DURING USE A .014 6.0 X 15 X 142CM AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT 10 ATMOSPHERES (ATM) DURING ITS INITIAL INFLATION. THEREFORE, IT WAS REPLACED WITH A SAME SIZE NEW NON-CORDIS BALLOON CATHETER AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT INJURY. THE LESION WAS THE LEFT RENAL ARTERY WITH STRONG CALCIFICATION AND SHORT STENOSIS. A .070 RENAL CURVE I VISTA BRITE TIP GUIDING CATHETER AND A 6F 25CM UNKNOWN SHEATH WERE USED DURING A RIGHT FEMORAL APPROACH. A NON-CORDIS GUIDEWIRE CROSSED THE LESION. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION AS IT WAS DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11342550 ↗MalfunctionDec 22, 2020Difficult to Remove; Difficult to Remove
AS REPORTED, A .014¿ 6.0 X 15 X 142CM AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WAS USED FOR PRE-PTA; BUT IT WAS NOT ABLE TO CROSS THE LESION. AFTER EXPANDING THE PART WHERE THE BALLOON CATHETER ENTERED THE STENOSIS A LITTLE AND DEFLATING; THE PHYSICIAN TRIED TO PASS THE DEVICE THROUGH THE STENOSIS AGAIN, BUT THE SAME ISSUE CONTINUED. THE PHYSICIAN THEN REMOVED THE BALLOON CATHETER, HOWEVER, IT GOT CAUGHT IN THE NON-CORDIS GUIDING SHEATH AND BECAME DIFFICULT TO BE REMOVED. FORTUNATELY, THE AVIATOR PLUS PTA BALLOON COULD BE REMOVED AND THE PROCEDURE RESUMED. THE DEVICE WAS REPLACED WITH AN UNKNOWN 4MM X 2CM BALLOON CATHETER, AND PTA WAS PERFORMED. AN UNKNOWN 6MM X 18MM STENT WAS USED AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT INJURY. THE TARGET LESION WAS AT THE ORIGIN PART OF THE RIGHT RENAL ARTERY. THE LESION WAS MODERATELY TORTUOUS WITH MODERATE CALCIFICATION. THE LESION HAD 90% STENOSIS AND WAS NOT A CTO. THE DEVICE WAS OPENED IN A STERILE FIELD. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, OR REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINK
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11055542 ↗MalfunctionNov 16, 2020Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, A .014" 6MM X 15MM X 142MM AVIATOR PLUS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WAS USED AS A PRE-DILATION; BUT IT WAS NOT ABLE TO BE INFLATED WHEN PRESSURE EXCEEDED ITS NOMINAL PRESSURE. THE DOCTOR CONFIRMED THAT THE BLOOD WAS PULLED OUT AND CONFIRMED THE RUPTURE. THERE WAS NO REPORTED PATIENT INJURY. THIS WAS A PERCUTANEOUS TRANSLUMINAL RENAL ANGIOPLASTY (PTRA) CASE. THE LESION WAS THE STENOSIS AT THE RIGHT RENAL ARTERY. AN APPROACH WAS MADE FROM THE RIGHT FEMORAL ARTERY. THE DEVICE WAS STORED IN THE CATH LAB FOR SIX HOURS. THE DEVICE WAS STORED, HANDLED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THE DEVICE WAS PREPPED NORMALLY BY MAINTAINING NEGATIVE PRESSURE. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, OR REMOVING THE PROTECTIVE BALLOON COVER. THERE WAS NO DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS, OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE BALLOON CATHETER WAS REMOVED EASILY AND INTACT. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS DUE TO INFECTIOUS DISEASE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10844554 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page