Skip to content

S.M.A.R.T. C09080MBRecalls & reported problems

Cordis US Corp. · PKG ASSY 9X080 SMART BIL 120CM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032024478), received between Feb 24, 2020 and Feb 24, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K042969); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Difficult or Delayed Positioning1
  • Difficult to Remove1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 24, 2020Difficult or Delayed Positioning; Difficult to Remove; Difficult or Delayed Positioning; Difficult to Remove

    AS REPORTED, THE DEPLOYMENT SYSTEM OF A 120CM X 9MM X 80 SMART CONTROL BILIARY STENT DID NOT ENGAGE AT ALL. THEREFORE, THE DEVICE WAS PULLED, HOWEVER, THE STENT GOT DISLODGED AND STUCK AT THE END OF THE SHEATH AND A CUTDOWN WAS PERFORMED TO RETRIEVE THE STENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9747253 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042969. One submission often covers many models, so check the product and lot details in each record.

  • Z-1508-2012 · Apr 9, 2012 · Terminated
    Product is labeled in part: Inner Label "***SMART Control Transhepatic Biliary Stent***REF Cat. No.***Lot No.***Use By***Cordis S.M.A.R.T. CONTROL Nitinol Stent Transhepatic Bilia
  • Z-1507-2012 · Apr 9, 2012 · Terminated
    Product is labeled in part: Inner Label "***S.M.A.R.T. Transhepatic Biliary Stent***REF Cat. No.***Lot No.***Use By***Cordis S.M.A.R.T. Nitinol Stent Transhepatic Biliary System**

Other Stents, Drains And Dilators For The Biliary Ducts devices

All Stents, Drains And Dilators For The Biliary Ducts devices →

Sources and updates