PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System PG1910BPSRecalls & reported problems
Cordis US Corp. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032038666), received between Mar 3, 2017 and Feb 19, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K033394); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Difficult to Remove1
- Inflation Problem1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 19, 2019Inflation Problem; Device Dislodged or Dislocated; Inflation Problem; Device Dislodged or Dislocated
AS REPORTED, THE GENESIS LONG STENT DEPLOYMENT SYSTEM WOULD NOT INFLATE TO DEPLOY THE MOUNTED STENT AND THE STENT DISLODGED IN THE PATIENT. THERE WAS NO PATIENT INJURY. ANOTHER UNKNOWN BALLOON WAS USED TO DEPLOY THE STENT AS THE STENT REMAINED AT THE TARGET LESION AFTER IT DISLODGED. THE BALLOON LUMEN OF THE GENESIS LONG WOULD NOT WORK AS THERE WAS A BLOCKED. THEREFORE, THE BALLOON AND/OR THE STENT THAT WAS MOUNTED ON IT, COULD NOT BE INFLATED. WHICH RESULTED IN THE STENT DISMOUNTED PREMATURELY. THE OPERATOR HAD TO USE ANOTHER BALLOON TO INFLATE THE STENT IN THE VESSEL WHICH FORTUNATELY REMAINED AT ITS TARGET LOCATION AND DIDN¿T MOVE OR EMBOLIZED. ONCE THE CATHETER WAS REMOVED FROM THE BODY, THE OPERATOR TRIED TO INFLATE THE BALLOON AGAIN TO FIGURE OUT THE ISSUE. THE WIRE LUMEN OF THE CATHETER WAS OKAY. THE BALLOON LUMEN OF THE CATHETER WOULD NOT FUNCTION, IT WOULD NOT INFLATE AT ALL. BLOCKED BY SOMETHING, PREVENTING THE BALLOON TO INFLATE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8349721 ↗InjuryMar 3, 2017Difficult to Remove; Device Dislodged or Dislocated; Difficult to Remove; Device Dislodged or Dislocated
AS REPORTED BY THE CUSTOMER VIA TELEPHONE, THE PALMAZ GENESIS PRO WAS INFLATED BUT THE BALLOON COULD NOT EXPAND NOR THE STENT COULDN'T BE DEPLOYED. THERE WAS DIFFICULTY WITHDRAWING THE SYSTEM OUT OF THE PATIENT AND THE STENT DISLODGED FROM THE BALLOON INTO THE PATIENT WHILE WITHDRAWING. SNARING WAS NECESSARY TO RETRIEVE THE STENT, NO PATIENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6378237 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033394. One submission often covers many models, so check the product and lot details in each record.
- Z-1363-2022 · May 20, 2022 · Open, ClassifiedPALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.