PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System PG2990BPXRecalls & reported problems
Cordis US Corp. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20705032039809), received between Jul 28, 2020 and Mar 4, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K033394); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated5
- Loss of or Failure to Bond1
- Material Protrusion/Extrusion1
- Premature Activation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, THE STENT OF A 29MM X 90MM PALMAZ GENESIS ON OPTA PRO STENT DELIVERY SYSTEM (SDS) CAME OFF OF THE BALLOON INSIDE OF AN UNKNOWN CATHETER SHEATH INTRODUCER (CSI) PRIOR TO DEPLOYMENT. THERE WAS NO DIFFICULTY REMOVING THE DELIVERY SYSTEM FROM THE PATIENT, AND THE STENT WAS ABLE TO BE REMOVED BY REMOVING THE SHEATH FROM THE PATIENT. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A SUBCLAVIAN LESION. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM ITS PACKAGING. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THE STENT WAS NOT MANIPULATED PRIOR TO INSERTION. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18832459 ↗MalfunctionNov 24, 2021Loss of or Failure to Bond; Premature Activation; Device Dislodged or Dislocated; Material Protrusion/Extrusion; Loss of or Failure to Bond; Premature Activation; Device Dislodged or Dislocated; Material Protrusion/Extrusion
AS REPORTED, THE STENT OF A LONG GENESIS/PRO 29 X 90 BIL 135 STENT DELIVERY SYSTEM (SDS) FELL OFF THE BALLOON. WHEN THEY WERE INSPECTING THE STENT BEFORE INSERTING INTO THE BODY, THE DOCTOR SAW IT WAS NOT CRIMPED SECURELY ON THE BALLOON. THE DOCTOR TRIED TO MANUALLY CRIMP THE STENT ON THE BALLOON BY HAND AND ALSO TRIED TO ROLL THE STENT AND BALLOON ON THE TABLE TECHNIQUE TO GET THE STENT ON THE BALLOON, BUT WAS UNSUCCESSFUL. THE PHYSICIAN WAS NOT CONFIDENT IN USING THE DEVICE FOR THE PROCEDURE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER, SMALLER PALMAZ GENESIS STENT. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE NEVER WENT IN THE BODY. DEVICE PACKAGING LOOKED NORMAL AND DID NOT HAVE DAMAGE. THE PHYSICIAN IS TRAINED AND HAS USED THE DEVICE IN THE PAST. THE TARGET LESION WAS A RENAL ARTERY. THE PHYSICIAN NOTICED THE STENT WAS LOOSE ON THE BALLOON AND TRIED TO "CRIMP/ROLL" THE STENT BACK, BUT WAS UNSUCCESSFUL. THE DILATION CATHETER OF THE SDS WAS SUCCESSFULLY USED LATER IN THE PROCEDURE TO POST DILATE. THE DEVICE WILL BE RETURNED FOR ANALYSIS. ADDENDUM: PRODUCT ANALYSIS REVEALS THE STENT WAS OBSERVED PARTIALLY EXPANDED AND SOME STRUTS WERE OBSERVED BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12868030 ↗MalfunctionJun 3, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, A PALMAZ LONG GENESIS STENT ON .035" 29 X 90 X 135 OPTA PRO BALLOON EXPANDABLE STENT DELIVERY SYSTEM WAS PUT ON THE WIRE (UNKNOWN); HOWEVER, IT COMPLETELY MOVED OFF THE BALLOON AND IT SLID BACK AND FORTH. THE EVENT WAS NOTICED BEFORE IT GOT TO THE SHEATH (UNKNOWN). THEREFORE, A NEW STENT (UNKNOWN) WAS OPENED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION. THERE WAS NO DAMAGE NOTED ON THE PACKAGING. THE DEVICE WAS STORED ACCORDING TO THE IFU. DURING REMOVAL FROM THE INNER PACKAGING, THE TRAY WAS CAREFULLY EXTRACTED WITH THE STENT AND DELIVERY SYSTEM AND INTRODUCER TUBE. THE TRAY WAS HELD IN ONE HAND AND WITH THE OTHER HAND, AND THE STAFF GENTLY GRASPED THE HUB AND CAREFULLY REMOVED THE STENT/DELIVERY SYSTEM FROM THE TRAY. WHEN THE DEVICE WAS INSPECTED, IT WAS NOT ADHERED TO THE BALLOON, SO A NEW STENT WAS OPENED. THERE HAD BEEN NO REPOSITIONING OR HAND CRIMP. BEFORE ATTEMPTING TO LOAD THE DEVICE ONTO THE WIRE, THE DEVICE WAS PREPPED PER IFU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11935380 ↗MalfunctionFeb 8, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, WHEN THE BALLOON OF A .035¿ 29MM X 30MM X 135CM PALMAZ GENESIS TRANSHEPATIC BILIARY STENT ON OPTA PRO PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON EXPANDABLE STENT DELIVERY SYSTEM (SDS) CAME OUT OF THE 7F NON-CORDIS SHEATH, THERE WAS NO STENT ON IT. THE PHYSICIAN SEARCHED ALL VESSELS FROM HEAD TO TOE ON FLURO BUT COULD NOT FIND A DISLODGED STENT. HE ALSO CHECKED IF IT WAS IN THE ¿SHUTTLE SHEATH¿ BUT FOUND NOTHING. THERE WAS NO REPORTED PATIENT INJURY. THE BALLOON WAS REMOVED TO INSPECT IT TO SEE IF THE STENT WAS ON IT. AFTERWARDS, THE PHYSICIAN USED THE SAME BALLOON FOR ANGIOPLASTY THE LESION. THE LESION WAS THE RIGHT COMMON CAROTID WHICH WAS CALCIFIED, TORTUOUS, AT A BIFURCATE AND IN AN ACUTE ANGLE WITH NINETY-FIVE PERCENT STENOSIS. THE DEVICE WAS STORED AND HANDLED PER THE INSTRUCTIONS FOR USE. THERE WAS NO DIFFICULTY EXPERIENCED IN PREPPING THE DEVICE. THE ACCESS SITE WAS FEMORAL, AND A CONTRALATERAL APPROACH USED. A COMPETITOR¿S SHEATH AND AN AQUATRACK GUIDEWIRE WERE USED. THE DEVICE WAS STORED AS PER LABELING AND WAS OPENED IN A STERILE FIELD. THE DEVICE WILL BE RETURNED FOR EVALUATION. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11298405 ↗MalfunctionJul 28, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, AFTER BACKING OUT THE 29 X 90 X 135 PALMAZ LONG GENESIS TRANSHEPATIC BILIARY STENT, THE USERS NOTICED THE STENT LAYING ON THE DRAPE UNDEPLOYED. THEREFORE, THE USERS DEPLOYED A NON-CORDIS STENT. THERE WAS NO REPORTED PATIENT INJURY. INITIALLY, THE USERS WENT TO LOAD THE DEVICE ON AN UNKNOWN WIRE AND DEPLOYED IT, HOWEVER, AFTER DEPLOYMENT, THE USERS WERE NOT ABLE TO SEE THE STENT UNDER FLOUROSCOPY (FLURO). AFTER THE USERS DEPLOYED A NON-CORDIS STENT, THEY CHECKED THE PALMAZ DEVICE SINCE THEY NEVER HAD THIS HAPPENED. THE DOCTOR PUT THE STENT ONTO THE UNKNOWN BALLOON AND PARTIALLY DEPLOYED IT, CURIOUS TO KNOW IF IT WAS THE STENT OR NOT. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10334141 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033394. One submission often covers many models, so check the product and lot details in each record.
- Z-1363-2022 · May 20, 2022 · Open, ClassifiedPALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page