PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System PG3910BPSRecalls & reported problems
Cordis US Corp. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20705032039830), received between Mar 29, 2021 and Jan 26, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K033394); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated4
- Deflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 26, 2024Deflation Problem; Device Dislodged or Dislocated; Deflation Problem; Device Dislodged or Dislocated
AS REPORTED, THE STENT OF A 4MM X 18MM PALMAZ BLUE ON AVIATOR PLUS STENT DELIVERY SYSTEM (SDS) DISLODGED FROM THE BALLOON WITHIN THE RENAL ARTERY AND THE BALLOON WAS UNABLE TO BE DEFLATED. AS A RESULT, THE DELIVERY SYSTEM WAS INTENTIONALLY CUT, AND THE BALLOON WAS DEFLATED WITH A SYRINGE TO REMOVE ENTIRE DELIVERY SYSTEM FROM THE PATIENT AND. A PERIPHERAL RETRIEVABLE FILTER TRAP WAS THEN USED TO CAPTURE AND RETRIEVE THE DISLODGED STENT AND THE PROCEDURE WAS COMPLETED. THIS WAS DURING A PROCEDURE TO TREAT VERTEBRAL ARTERY STENOSIS. THE LESION HAD MODERATE CALCIFICATION, MILD TORTUOSITY, AND A 75% STENOSIS. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WERE NO DIFFICULTIES REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. A NON-CORDIS BALLOON CATHETER WAS USED TO PRE-DILATE THE LESION. THE AVIATOR PLUS SDS WAS NOT KINKED AT ANY TIME DURING THIS PROCEDURE. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18586545 ↗MalfunctionDec 7, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, WHEN REMOVING A 39MM X 10MM PALMAZ GENESIS ON OPTA PRO STENT DELIVERY SYSTEM (SDS), THE STENT FELL OFF OF THE BALLOON. THE STENT WAS NOT MANIPULATED AND FELL OFF OF THE DEVICE PRIOR TO PREPARATION. THE DEVICE NEVER WENT INTO THE PATIENT AND A NON-CORDIS STENT WAS USED TO TREAT THE INTENDED LESION. THERE WERE NO PATIENT INJURIES REPORTED. THE PALMAZ GENESIS STENT WAS TO BE IMPLANTED IN THE ILIAC ARTERY DURING AN ENDOVASCULAR ANEURYSM REPAIR (EVAR) PROCEDURE. THE PACKAGING WAS OPENED CORRECTLY, AND THE STENT DID NOT APPEAR PREMATURELY EXPANDED. THE DEVICE WAS DISCARDED AND WILL NOT BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18284450 ↗MalfunctionMay 12, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, THE STENT OF A 39MM X 10MM PALMAZ GENESIS BILIARY ON OPTA PRO .035 STENT DELIVERY SYSTEM (SDS) FELL OF DURING PREPARATION, PRIOR TO PATIENT USE. THERE WAS NO INJURY TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS UNAVAILABLE. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16919906 ↗InjuryMar 29, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
PHYSICIAN WAS CANNULATING THE ILIAC ARTERY, HOWEVER, WHEN TRYING TO PLACE A 32MM PALMAZ LONG GENESIS BILIARY STENT ON .035¿ 39 X 10 X 80 OPTA PRO PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER; THE STENT SLIDES OFF THE BALLOON PRIOR TO PROPER PLACEMENT AND PROPER DEPLOYMENT/INFLATION. PHYSICIAN WAS ABLE TO SAVE THE STENT USING VARIOUS TECHNIQUES. NO HARM WAS DONE TO THE PATIENT. THE PRODUCT WAS STORED AS PER LABELING AND OPENED IN A STERILE FIELD. THE ACCESS SITE WAS FEMORAL. THE LESION HAD LITTLE CALCIFICATION. THERE WAS LITTLE VESSEL TORTUOSITY. A 7F CORDIS BRITE TIP SHEATH WAS USED. A .035 NON-CORDIS GUIDEWIRE WAS USED. THERE WERE NO VISIBLE SIGNS OF DEVICE/PACKAGE DAMAGE PRIOR TO USE. THERE WERE NO ANOMALIES NOTED WHEN THE DEVICE WAS TAKEN OUT OF THE PACKAGE. THE DEVICE WAS PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY EXPERIENCED IN PREPPING THE DEVICE. THE PRODUCT, OR ANY OF THE OTHER DEVICES USED WITH IT, HAD NOT BEEN RESTERILIZED. DURING PREP, THE STENT WAS NOT MANIPULATED. THE STENT WAS PROPERLY MOUNTED ON THE SYSTEM WHEN INSPECTED PRIOR TO USE. PRIOR TO INSERTING THE SDS IN THE CATHETER, THE BALLOON WAS NOT INFLATED NOR PARTIALLY INFLAT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11582696 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033394. One submission often covers many models, so check the product and lot details in each record.
- Z-1363-2022 · May 20, 2022 · Open, ClassifiedPALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page