PALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System PG5980BPSRecalls & reported problems
Cordis US Corp. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032040393), received between Mar 22, 2021 and May 24, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K033394); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated2
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 24, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, DURING PREP, A PALMAZ GENESIS .035¿ 8X59 BILIARY STENT WAS OPENED AND FELL OFF ON THE BACK TABLE BEFORE INSERTING THROUGH THE SHEATH TO THE TREATMENT SITE. A NEW STENT WAS OPENED AND USED TO COMPLETED THE PROCEDURE SUCCESSFULLY. THERE WAS NO REPORTED PATIENT INJURY. THE INTENDED PROCEDURE WAS REPORTED AS AN ILIAC ARTERY STENTING. THE PRODUCT WAS STORED, AND PREPPED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DAMAGE NOTED TO THE PRODUCT PACKAGING UPON INSPECTION PRIOR TO USE AND NO DIFFICULTY REMOVING THE PRODUCT FROM THE PACKAGING. THE STENT WAS NOT CRIMPED USING A CRIMPING TOOL, OR HAND CRIMP. THE BALLOON WAS NOT PARTIALLY INFLATED TO MAINTAIN THE STENT IN PLACE. THERE WAS NO DAMAGE OR DEFORMATION NOTED ON THE STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16991769 ↗InjuryMar 22, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED, THE 8MM X 59MM X 80CM PALMAZ GENESIS TRANSHEPATIC BILIARY STENT ON .035¿ OPTA PRO PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) DELIVERY SYSTEM DISLODGED FROM THE BALLOON INSIDE THE PATIENT. THE PATIENT WAS FINE, BUT THEY HAD TO USE ANOTHER UNKNOWN STENT. THE LEFT ILIAC ARTERY WAS PRE DILATED WITH A POWERFLEX 5 X 40 AT 10 ATMOSPHERES (ATM). THEY CROSSED THE AREA WITH A 9 X 39 GENESIS WITH NO ISSUES PRIOR TO PLACING THE 8 X 59. THE PATIENT HAD A MEDICAL HISTORY OF PERIPHERAL VASCULAR DISEASE. THE INTENDED PROCEDURE/TARGET LESION WAS THE ILIAC ARTERY . VESSEL / LESION CHARACTERISTICS, THE LESION HAD LITTLE CALCIFICATION. THE VESSEL HAD LITTLE TORTUOSITY . THE VESSEL HAD NO ANGULATION. THE DEVICE WAS NOT USED FOR A CHRONIC TOTAL OCCLUSION (TOTAL OCCLUSION >3 MONTHS). THE DEVICE WAS NOT RESTERILIZED. THERE WERE NO ANOMALIES NOTED WHEN REMOVED FROM THE PACKAGE. THERE WERE NO ANOMALIES NOTED DURING PREP. THE DEVICE WAS NOT INSERTED THROUGH A STOPCOCK INSTEAD OF A HEMOSTATIC VALVE. THE SEGMENT WAS LEFT I THE ILIAC ARTERY. THE PATIENT IS STABLE AFTER THE EVENT. THE DEVICE WILL NOT BE RETURNED AS IT REMAINED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 11544683 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033394. One submission often covers many models, so check the product and lot details in each record.
- Z-1363-2022 · May 20, 2022 · Open, ClassifiedPALMAZ GENESIS Transhepatic Biliary Stent on OPTA PRO .035" Delivery System, Catalog numbers PG2990BPS, PG2990BPX, PG3990BPS, PG3990BPX.
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page