POWERFLEX 4400708XRecalls & reported problems
Cordis Corporation · POWERFLEXPRO 7MM8CM 135 · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032059746), received between Aug 8, 2023 and Aug 8, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2023Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, A 7MM X 8CM 135CM POWERFLEX PRO PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WAS USED SUCCESSFULLY AND WAS REMOVED TO BE USED LATER IN THE PROCEDURE. HOWEVER, UPON FLUSHING THE BALLOON CATHETER PRIOR TO RE-USING IT, THE CATHETER SHAFT BURST. THE PROCEDURE WAS ENDED AND THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS OCCURRED DURING A PROCEDURE IN WHICH A LOWER LEG ARTERIOGRAM WAS PERFORMED. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND WAS ABLE TO MAINTAIN NEGATIVE PRESSURE DURING PREPARATION. THERE WERE NO DIFFICULTIES REMOVING THE STERILE PACKAGING COMPONENTS. THE INITIAL INFLATION OF THE BALLOON WAS BELOW THE RATED BURST PRESSURE (RBP). AFTER INITIAL USE OF THE DEVICE, THE DEVICE WAS COILED AND LAID FLAT ON THE SCRUB TABLE. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17483347 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.