POWERFLEX 4401008XRecalls & reported problems
Cordis Corporation · POWERFLEXPRO 10MM8CM 135 · Class 2
All models of POWERFLEX (34)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032060100), received between Nov 1, 2021 and Apr 13, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
- Detachment of Device or Device Component1
- Difficult to Remove1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 13, 2022Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component; Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component
AS REPORTED, THE BALLOON OF A 10MM X 8CM 135CM POWERFLEX PRO .035 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED DURING AN INFLATION TO 12 ATMOSPHERES (ATM) IN THE ILIAC VEIN. AFTER THE RUPTURE, THE POWERFLEX PRO WAS DIFFICULT TO REMOVE. TO REMOVE THE POWERFLEX PRO PTA BALLOON CATHETER, TRANSECTION OF THE DEVICE WAS REQUIRED, AND AN INCISION AT THE PUNCTURE SITE WAS MADE. ONCE THE BALLOON WAS REMOVED, THE PATIENT WAS OBSERVED UNTIL STABILIZATION AND THE PROCEDURE WAS ENDED. THE POWERFLEX PRO WAS BEING USED TO TREAT ILIAC VEIN IN-STENT RESTENOSIS WHICH WAS NOT REPORTED AS A CHRONIC TOTAL OCCLUSION (CTO). THE ILIAC VEIN HAD A 75% PERCENT STENOSIS AND SHOWED NO SIGNS OF CALCIFICATION OR TORTUOSITY. THE DEVICE WAS STORED AND PREPARED PER THE INSTRUCTIONS FOR USE (IFU) AND MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, THE PROTECTIVE BALLOON COVER, OR ANY OF THE STERILE PACKAGING COMPONENTS. HENGRUI CONTRAST WAS USED, AND THE DEVICE WAS PREPARED AT A 1:1 CONTRAST TO SALINE RATIO. THERE WAS NO RESISTANCE OR FRICTION EXPERIENCED WHILE THE BALLOON WAS BEING INSERTED INTO THE PATIENT, THE BALLOON WAS ABLE T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14100973 ↗MalfunctionNov 1, 2021Leak/Splash; Leak/Splash
DURING THE PURGING OF A 10MM X 8CM 135 POWERFLEX PRO PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER ON THE TABLE, IT WAS OBSERVED THAT A JET OF WATER CAME OUT OF THE SHAFT OF THE BALLOON. THEREFORE, IT WAS NOT USED IN THE PATIENT AND A NEW POWERFLEX BALLOON (UNKNOWN) WAS USED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED AS PER THE INSTRUCTIONS FOR USE (IFU) AND WAS OPEN IN A STERILE FIELD. THE INTENDED PROCEDURE WAS A BILIARY ANGIOPLASTY. THE LESION WAS NOT CALCIFIED. THERE VESSEL WAS SLIGHTLY TORTUOUS. THE PERCENTAGE OF STENOSIS WAS MILD. THE DEVICE WAS NOT USED FOR A CHRONIC TOTAL OCCLUSION (TOTAL OCCLUSION >3 MONTHS). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, THE PROTECTIVE BALLOON COVER, OR THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THE BALLOON CATHETER DID NOT KINK WHILE BEING USED. THE DEVICE WILL BE RETURNED FOR EVALUATION. PHOTOGRAPHS WERE PROVIDED AND AVAILABLE FOR REVIEW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12728216 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page