SMART FLEX SF06100MBRecalls & reported problems
Cordis US Corp. · SMART FLEX 6x100 BIL, 120cm · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032065648), received between Oct 23, 2023 and Oct 23, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Difficult or Delayed Positioning” (1 report).
1 FDA recall record name this device's identifier.
6 other recalls cover devices cleared under the same FDA submission (K121125); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 15, 2017Terminated
Potential for cracked luer hubs
Z-0324-2018 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Difficult or Delayed Positioning1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 23, 2023Difficult or Delayed Positioning; Difficult or Delayed Positioning
AS REPORTED, DURING DEPLOYMENT OF A 6MM X 100MM 120CM SMART FLEX BILIARY SELF-EXPANDING STENT (SES) DELIVERY SYSTEM, THE PHYSICIAN FULLY LOOSENED THE TOUHY BORST VALVE TO INITIATE STENT DEPLOYMENT. THEIR HANDS WERE PLACED ON THE TOUHY VALVE AND THE METAL PUSH ROD, AND THE STENT DEPLOYMENT WAS PERFORMED VIA THE PIN AND PULL METHOD. THE USER HELD THE HANDLE OF THE DELIVERY SYSTEM FLAT AND STRAIGHT OUTSIDE OF THE PATIENT DURING ATTEMPTED DEPLOYMENT. HOWEVER, RESISTANCE WAS FELT, AND THE PHYSICIAN HAD TO INCREASE THE FORCE OF THE PIN AND PULL WHICH CAUSED THE Y-ADAPTOR TO DISLODGE FROM THE OUTER LAYER OF THE SDS. THE DETACHMENT OF THE TOUHY VALVE MADE THE DEPLOYMENT OF THE STENT DIFFICULT TO PERFORM AND THE STENT WAS PARTIALLY DEPLOYED. THE PHYSICIAN WAS ADVISED TO PULL BACK ON THE SDS TO CONTINUE DEPLOYMENT; HOWEVER, THIS CAUSED THE PARTIALLY DEPLOYED SDS TO MIGRATE PROXIMALLY. THE PHYSICIAN ATTEMPTED TO REMOVE THE PARTIALLY DEPLOYED STENT THROUGH A NON-CORDIS CATHETER SHEATH INTRODUCER (CSI) WHILE THE STENT WAS STILL ATTACHED TO THE DELIVERY SYSTEM; HOWEVER, THE REMOVAL WAS UNSUCCESSFUL, AND THE STENT WAS DEPLOYED PROXIMALLY TO ITS INTENDED LOCATION. AS A RESULT, A NEW 6MM X 120MM SM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17984743 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121125. One submission often covers many models, so check the product and lot details in each record.
- Z-0216-2022 · Oct 1, 2021 · Open, ClassifiedSMART FLEX 9x60 BIL, 80cm The S.M.A.R.T. Flex Biliary Stent System is indicated for use in the palliation of malignant strictures in the biliary tree.
- Z-0220-2022 · Oct 1, 2021 · Open, ClassifiedSMART FLEX 10x100 BIL, 80cm The S.M.A.R.T. Flex Biliary Stent System is indicated for use in the palliation of malignant strictures in the biliary tree.
- Z-0218-2022 · Oct 1, 2021 · Open, ClassifiedSMART FLEX 6x150 BIL, 120cm The S.M.A.R.T. Flex Biliary Stent System is indicated for use in the palliation of malignant strictures in the biliary tree.
- Z-0219-2022 · Oct 1, 2021 · Open, ClassifiedSMART FLEX 8x60 BIL, 80cm The S.M.A.R.T. Flex Biliary Stent System is indicated for use in the palliation of malignant strictures in the biliary tree.
- Z-0217-2022 · Oct 1, 2021 · Open, ClassifiedSMART FLEX 6x120 BIL, 120cm The S.M.A.R.T. Flex Biliary Stent System is indicated for use in the palliation of malignant strictures in the biliary tree.
- Z-0221-2022 · Oct 1, 2021 · Open, ClassifiedSMART FLEX 10x100 BIL, 120cm The S.M.A.R.T. Flex Biliary Stent System is indicated for use in the palliation of malignant strictures in the biliary tree.
Other Stents, drains and dilators for the biliary ducts devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.