Skip to content

SABER 48002025XRecalls & reported problems

Cordis US Corp. · SABER 2MM25CM 150 · Class 2

All models of SABER (51)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20705032069226), received between Apr 16, 2021 and May 4, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

1 other recall cover devices cleared under the same FDA submission (K133843); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Difficult to Remove1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 4, 2025Difficult to Remove; Difficult to Remove

    AS REPORTED, THE 2MM X 250MM 150CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) DILATATION CATHETER WAS USED TO PRE-DILATE AND THE BALLOON WAS NOT ABLE TO BE INFLATED. THERE WAS A LITTLE RESISTANCE FOUND WHEN REMOVING THE DEVICE FROM THE PATIENT, BUT IT WAS REMOVED INTACT (IN ONE PIECE) AND REPLACED WITH ANOTHER BALLOON AND WAS USED WITH NO PROBLEM. THERE WERE NO REPORTS OF PATIENT INJURY. THE DEVICE WAS USED TO TREAT A LOWER EXTREMITY OCCLUSION. THERE WERE NO DAMAGES FOUND ON THE DEVICE OR DEVICE PACKAGING PRIOR TO OPENING. THE DEVICE WAS STORED AND PREPPED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PROTECTIVE BALLOON COVER OR ANY OF THE STERILE PACKAGING COMPONENTS. THE TARGET SITE WAS 90% STENOSED. THE ACCESS SITE WAS NORMAL. THE DEVICE DID NOT MAINTAIN NEGATIVE PRESSURE DURING PREPARATION. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND AT ANY POINT DURING THE PROCEDURE. THE DEVICE WAS ABLE TO CROSS THE LESION AND DID NOT KINK AT ANY TIME DURING THE PROCEDURE. THERE WERE NO REPORTS OF PATIENT INJURY. THE BALLOON DID NOT LEAK FROM ANY SEGMENT. THE USER WAS TRAINED IN THE USE OF THE DEVICE. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21957559 ↗
  • MalfunctionFeb 2, 2024Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, THE PATIENT WAS IN THE ENDOCRINOLOGY DEPARTMENT WITH A DIABETIC FOOT, HAVING BALLOON DILATATION WAS PROPOSED. SURGICAL DILATATION WAS PERFORMED USING 2MM 25CM SABER; HOWEVER, THE BALLOON WAS FOUND TO BE RUPTURED AND THE CONTRAST AGENT OOZED OUT. THE BALLOON WAS REPLACED AND DILATATION WAS PERFORMED WITHOUT ANY ABNORMALITY IN THE PATIENT. THERE WAS NO REPORTED INJURY TO THE PATIENT. ADDITIONAL INFORMATION AND DEVICE RETURN WAS REQUESTED; HOWEVER, BOTH WERE NOT OBTAINED AFTER MULTIPLE ATTEMPTS MADE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18634349 ↗
  • MalfunctionApr 16, 2021Leak/Splash; Leak/Splash

    AS REPORTED, THE BALLOON OF A 2MM X 25CM 150 SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER HAD LEAKAGE AND FAILED TO BE INFLATED. THERE WAS NO REPORTED PATIENT INJURY. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, WHEN REMOVING THE PROTECTIVE BALLOON COVER OR WHEN REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE DEVICE PREPPED NORMALLY. THE CONTRAST TO SALINE RATIO WAS 1:1. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE. VESSEL /LESION CHARACTERISTICS: THE LESION HAD CALCIFICATION, NO VESSEL TORTUOSITY. , THE VESSEL WAS 90% STENOSED . THE DEVICE WAS USED FOR A CHRONIC TOTAL OCCLUSION (TOTAL OCCLUSION >3 MONTHS). THERE WAS NO DIFFICULTY ADVANCING THE BALLOON CATHETER THROUGH THE VESSEL. THERE WAS NO DIFFICULTY CROSSING THE LESION. THE CATHETER WAS NEVER IN ACUTE BEND. THE PLUNGER DID DEPRESS INTO THE SYRINGE/INDEFLATOR WHEN TRYING TO INFLATE. THE USER WAS ABLE TO DEPRESS THE PLUNGER INTO THE SYRINGE/INDEFLATOR WHEN TRYING TO INFLATE. THE MAXIMUM INFLATION PRESSURE WAS ZERO. THE DEVICE WIL

    Patient problems: Insufficient Information

    MAUDE report 11678682 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133843. One submission often covers many models, so check the product and lot details in each record.

  • Z-2235-2021 · Jul 19, 2021 · Terminated
    Cordis SABER PTA Balloon Dilation Catheter, 4F, REF 48004010X

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page