SABER 48002506XRecalls & reported problems

Cordis US Corp. · SABER 2.5MM6CM 150 · Class 2

All models of SABER (51)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20705032069264), received between May 3, 2021 and Nov 6, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

1 other recall cover devices cleared under the same FDA submission (K133843); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Leak/Splash1
  • Partial Blockage1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2023Partial Blockage; Partial Blockage

    AS REPORTED, A 2.5MM X 6CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) DILATATION CATHETER WOULD NOT FLUSH, AND THE GUIDEWIRE WOULD NOT PASS THROUGH THE BALLOON WIRE PORT. THE BALLOON WAS PARTIALLY LOADED. THE ISSUE OCCURRED DURING PREP. A NON-CORDIS BALLOON WAS USED AS REPLACEMENT. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS BEING USED FOR PERIPHERAL VASCULAR INTERVENTION. THE DEVICE WAS STORED, AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP OR REMOVING THE PROTECTIVE BALLOON COVER NOR DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THE DEVICE DID NOT KINK AT ANY TIME DURING USE. THE WIRE WAS WIPED WITH A WET TELFA/GAUZE AFTER THE WIRE WAS REMOVED AND BEFORE THE WIRE WAS LOADED ON. IT IS NOT KNOWN IF THE OBSTRUCTION WAS DUE TO MATERIAL THAT COULD POSSIBLY BE INJECTED INTO THE PATIENT. THE DEVICE WAS RETURNED FOR EVALUATION. THE PRODUCT EVALUATION IDENTIFIED POLYETHYLENE PARTICLES AS THE CAUSE OF THE OBSTRUCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18075328 ↗
  • MalfunctionSep 8, 2022Leak/Splash; Leak/Splash

    AS REPORTED, DURING AN INFLATION TO 10 ATMOSPHERES (ATM) OF A 2.5MM X 6CM 150M SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER, THERE WAS A LEAKAGE COMING FROM THE BALLOON. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A LESION IN THE ANTERIOR TIBIAL ARTERY. THE LESION WAS NOT A CHRONIC TOTAL OCCLUSION (CTO) AND THERE WAS NO CALCIFICATION OR VESSEL TORTUOSITY PRESENT. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. A 50/50 CONTRAST AND SALINE MIXTURE WAS USED TO PREP THE BALLOON. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, THE STYLET, OR ANY OF THE STERILE PACKAGING COMPONENTS AND THE DEVICE WAS NOT KINKED OR DAMAGED IN ANY WAY PRIOR TO BEING INSERTED INTO THE PATIENT. THERE WAS NO RESISTANCE EXPERIENCED DURING INSERTION OF THE DEVICE. THE DEVICE WAS ABLE TO TRACK TOWARD THE LESION AND CROSS THE LESION WITHOUT DIFFICULTY. THE DEVICE WAS NEVER PLACED IN AN ACUTE BEND, AND IT DID NOT KINK DURING ITS USE. THE DEVICE WAS ABLE TO BE REMOVED EASILY AND REMAINED IN ONE PIECE DURING ITS REMOVAL. THE DEVICE WAS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15382461 ↗
  • MalfunctionMay 3, 2021Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, A 2.5MM X 6CM 150 SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED ¿WITHOUT HITTING THE BLASTING PRESSURE¿. IT WAS THEN WITHDRAWN. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11759220 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133843. One submission often covers many models, so check the product and lot details in each record.

  • Z-2235-2021 · Jul 19, 2021 · Terminated
    Cordis SABER PTA Balloon Dilation Catheter, 4F, REF 48004010X

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →