SABER 48002530XRecalls & reported problems
Cordis Corporation · SABER 2.5MM30CM 150 · Class 2
All models of SABER (40)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 20705032069295), received between Oct 21, 2019 and Oct 17, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K133843); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel2
- Leak/Splash2
- Material Puncture/Hole2
- Detachment of Device or Device Component1
- Difficult to Open or Remove Packaging Material1
- Material Frayed1
- Material Split, Cut or Torn1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, AN UNKNOWN GUIDEWIRE FRACTURED WHILE USING A 2.5MM X 30CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER. THERE WAS NO REPORTED INJURY TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. THE DEVICE WILL BE RETURNED FOR EVALUATION. ADDENDUM: THE DEVICE WAS RETURNED SEPARATED AT THE BALLOON PORTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23321937 ↗MalfunctionAug 29, 2025Material Frayed; Material Split, Cut or Torn; Material Frayed; Material Split, Cut or Torn
AS REPORTED, THE 2.5MM 30CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON WAS INTENDED TO DILATE A STENOTIC LESION, BUT AFTER ADVANCING TO A CERTAIN POSITION, IT COULD NO LONGER BE PUSHED FORWARD. UPON WITHDRAWAL OF THE SYSTEM, IT WAS FOUND THAT THE BALLOON SHAFT HAD KINKED. THE CASE WAS COMPLETED BY WITHDRAWING THE DEVICE AND USING A NON-CORDIS BALLOON. THE PRODUCT EVALUATION OBSERVED THAT THE BODY/SHAFT WAS TORN. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCT WAS STORED AND INSPECTED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). NO PACKAGING DAMAGE OR DIFFICULTY REMOVING THE PRODUCT FROM THE PACKAGING OR HOOP WAS REPORTED. THE DEVICE APPEARED NORMAL PRIOR TO USE, AND NO KINKS OR DAMAGE WERE OBSERVED BEFORE INSERTION. A NON-CORDIS .018 GUIDEWIRE WAS USED AND SUCCESSFULLY CROSSED THE LESION. THE TARGET LESION WAS A MODERATE STENOSIS LOCATED BELOW THE KNEE, WITH MODERATE CALCIFICATION AND MODERATE TORTUOSITY. KINKS OR OTHER DAMAGE WERE NOTED AFTER THE DEVICE WAS REMOVED FROM THE PATIENT. THE DEVICE WAS NOT USED FOR A CHRONIC TOTAL OCCLUSION. NO SIGNIFICANT RESISTANCE WAS ENCOUNTERED DURING THE PROCESS. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22922144 ↗MalfunctionMay 27, 2025Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material; Tear, Rip or Hole in Device Packaging; Difficult to Open or Remove Packaging Material
AS REPORTED, THE SURGEON CHOSE A 2.5-300 MM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON FOR DILATION. UPON OPENING THE OUTER PACKAGING, IT WAS FOUND THAT THE STERILE PACKAGING BAG WAS ALREADY OPEN. THEREFORE, A NEW 2.5 BALLOON WAS USED FOR THE SURGERY. THERE WAS NO REPORTED INJURY TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED; HOWEVER, THE INFORMATION WAS NOT OBTAINED AFTER MULTIPLE ATTEMPTS. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22087523 ↗MalfunctionApr 30, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, DURING 2.5MM 30CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON INFLATION, THE BALLOON WAS FOUND TO BE UNABLE TO FILL. AN EXAMINATION OUTSIDE THE PATIENT REVEALED THE BALLOON WAS RUPTURED. THE BALLOON RUPTURED BELOW THE RATED BURST PRESSURE (RBP). THE PROCEDURE WAS COMPLETED BY SWITCHING TO ANOTHER BALLOON. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCT WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, THE PROTECTIVE BALLOON COVER, OR THE STYLET AND OTHER STERILE PACKAGING COMPONENTS. NO KINKS OR OTHER DAMAGES WERE NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE DEVICE WAS PREPPED PER THE IFU AND MAINTAINED NEGATIVE PRESSURE DURING PREP. THE INTENDED PROCEDURE WAS PRE-DILATATION OF THE SUPERFICIAL FEMORAL ARTERY. THE LESION WAS SEVERELY CALCIFIED. THE VESSEL HAD NO TORTUOSITY. THE PERCENTAGE OF STENOSIS WAS 80%. THE DEVICE WAS USED FOR A CHRONIC TOTAL OCCLUSION (CTO). NO RESISTANCE OR FRICTION WAS ENCOUNTERED WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE. NO GUIDE CATHETER WAS USED. THERE WAS DIFFICULTY CROSSING THE LESION WITH THE BALLOON CAT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21935396 ↗MalfunctionJan 20, 2023Leak/Splash; Leak/Splash
AS REPORTED, AFTER INFLATION OF THE SABER 2.5MM 30CM BALLOON TO 10ATM, IT WAS NOTICED THAT THE BALLOON WASN¿T HOLDING ITS SIZE. IT WAS NOTED TO BE SMALLER VIA FLUOROSCOPY, THOUGH IT WASN¿T BEING DEFLATED. THE PHYSICIAN REMOVED BALLOON FROM PATIENT AND INFLATED IT OUTSIDE OF THE BODY TO CHECK FOR LEAKS. THE DEVICE WAS REMOVED INTACT FROM THE PATIENT. IT WAS OBSERVED BY MD, THE TECH, AND SALES REP, THAT THERE WAS LEAKING NOTED TOWARD THE PROXIMAL END OF THE BALLOON. THE CASE WAS COMPLETED WITH THE USE OF A NON-CORDIS BALLOON. NO HARM CAME TO PATIENT. THE PROCEDURE BEING PERFORMED WAS A LEFT LEG ANGIOGRAM, WITH ANTEGRADE APPROACH. A 5F NON-CORDIS SHEATH WAS BEING USED. THE DEVICE WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, STYLET, OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE DEVICE WAS PREPPED PER THE IFU AND WAS ABLE TO MAINTAIN NEGATIVE PRESSURE. THE INTENDED LESION WAS IN THE LEFT POSTERIOR TIBIAL ARTERY. THE LESION WAS NOTED TO HAVE NO CALCIFICATION AND NO TORTUOSITY. T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16215354 ↗MalfunctionDec 13, 2021Leak/Splash; Leak/Splash
AS REPORTED, THE BALLOON ON THE SABER 2.5MM X 30CM CATHETER WAS NOTED TO HAVE LEAKAGE FROM THE MIDDLE SECTION OF THE BALLOON. THIS WAS NOTED AFTER THE DEVICE WAS INSERTED INTO THE VESSEL AND COULD NOT BE FILLED, AFTER THIS THE BALLOON WAS REMOVED FROM FROM THE PATIENT. THE CASE WAS COMPLETED WITH A NEW CORDIS BALLOON (SAME PRODUCT). THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCT WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, OR REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS, OR ANY DAMAGES NOTED ON THE DEVICE PRIOR TO INSERTING INTO THE PATIENT. THE DEVICE WAS PREPPED NORMALLY AND WAS ABLE TO MAINTAIN NEGATIVE PRESSURE. THE TARGET LESION WAS IN THE LOWER EXTREMITY WITH ARTERIOSCLEROSIS. THE LESION WAS NOTED TO HAVE NO CALCIFICATION OR TORTUOSITY; HOWEVER, THE LESION WAS NOTED TO HAVE A 75% STENOSIS. THE DEVICE WAS NOT BEING USED FOR A CHRONIC TOTAL OCCLUSION. THERE WAS NO RESISTANCE/FRICTION WHILE THE BALLOON CATHETER WAS INSERTED INTO THE ROTATING HEMOSTATIC VALVE, GUIDE CATHETER, OR WHILE ADVANCING THROUGH THE VESSEL. T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12984662 ↗MalfunctionOct 19, 2020Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, THEY NOTICED A HOLE PROXIMAL 5CM BEFORE THE BALLOON OF A 2.5MM X 30CM X 150CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER. THEY NOTICED THIS WHEN THEY WERE GOING TO WET IT WITH COMPRESSION AS THEY ALWAYS DO, SO THEY DID NOT HAVE THE UNKNOWN WIRE INSIDE YET. THE PRODUCT WAS NEVER ENTERED INSIDE THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY. THE PRODUCT WAS NEVER USED INSIDE THE PATIENT. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, FROM THE PROTECTIVE BALLOON COVER OR WHEN REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT THE PRODUCT INTO THE PATIENT. THE DEVICE WAS OPENED IN A STERILE FIELD. THE USER WAS TRAINED WITH THE DEVICE. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10697307 ↗MalfunctionSep 1, 2020Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, THE 2.5MM X 30CM X 150CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER (BC) HAS A DEFECT ON THE SHAFT. THERE WAS A HOLE IN THE OUTER LAYER ON THE DEVICE. THE DEFECT WAS NOTED WHEN THE INNER PACKAGE WAS OPENED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS NOT USED FOR A PATIENT. THE DEVICE WAS OPENED IN STERILE FIELD. THE DEVICE WAS STORED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE PACKAGING. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP OR REMOVING THE PROTECTIVE BALLOON COVER. THE DEVICE WILL BE RETURNED FOR ANALYSIS. IMAGE IS AVAILABLE FOR REVIEW.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10476039 ↗MalfunctionOct 21, 2019Burst Container or Vessel; Burst Container or Vessel
DURING THE PROCEDURE, THE DOCTOR OBSERVED A HOLE IN THE SABER BALLOON CATHETER WHEN A LEAK APPEARED DURING PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA). THERE WAS NO PATIENT INJURY. A COMPETITOR'S 0.018" GUIDE WIRE AND 6F SHEATH WAS USED. THE DEVICE IS AVAILABLE FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9215672 ↗
FDA clearance and classification
Cleared through 510(k) K133843 on Jun 27, 2014 (decision: Substantially Equivalent). Submitted by Cordis Corporation as “Saber Percutaneous Transluminal Angioplasty Dilatation Catheter”. The FDA review took 191 days. FDA record ↗
The FDA regulates product code LIT (catheter, angioplasty, peripheral, transluminal) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K133843. One submission often covers many models, so check the product and lot details in each record.
- Z-2235-2021 · Jul 19, 2021 · TerminatedCordis SABER PTA Balloon Dilation Catheter, 4F, REF 48004010X
Common questions about the SABER 48002530X
Has the SABER 48002530X been recalled?
No FDA recall record in our data names this exact device identifier (DI 20705032069295), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the SABER 48002530X?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Oct 21, 2019 and Oct 17, 2025. In those reports, the most common specific device problems are “Burst Container or Vessel” (2), “Leak/Splash” (2) and “Material Puncture/Hole” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SABER 48002530X?
None of the reports are classified as deaths, 0 as injuries and 9 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SABER 48002530X FDA approved?
It was cleared through 510(k) K133843 on Jun 27, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the SABER 48002530X, and is it still sold?
The FDA's device database lists Cordis Corporation as the company responsible for it (brand SABER, model 48002530X). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal880 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal439 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal383 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal280 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal128 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page