SABER 48005025XRecalls & reported problems
Cordis Corporation · SABER 5MM25CM 150 · Class 2
All models of SABER (40)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 20705032069561), received between Sep 9, 2020 and Jul 8, 2025 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (6 reports).
1 other recall cover devices cleared under the same FDA submission (K133843); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel6
- Deflation Problem1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2025Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE 5MM 25CM .018 SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED. THE PROCEDURE WAS COMPLETED USING ANOTHER .018 SABER BALLOON. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE DEVICE HAD BEEN STORED AND PREPARED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP. THERE WAS NO DIFFICULTY REMOVING THE PROTECTIVE BALLOON COVER. THERE WAS NO DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. NO KINKS OR OTHER DAMAGES WERE NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE DEVICE WAS PREPPED PER THE IFU. THE DEVICE PREPPED NORMALLY, MAINTAINING NEGATIVE PRESSURE. THE INTENDED PROCEDURE WAS FOR THE SUPERFICIAL FEMORAL ARTERY (SFA). THE LESION WAS SEVERELY CALCIFIED. THE VESSEL TORTUOSITY WAS SEVERE. THERE WAS NO RESISTANCE OR FRICTION WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE. THERE WAS NO RESISTANCE OR FRICTION WHILE INSERTING THE BALLOON THROUGH THE GUIDE CATHETER. THERE WAS NO DIFFICULTY ADVANCING THE BALLOON CATHETER THROUGH THE VESSEL. THERE WAS NO DIFFICULTY CROSSING THE LESION WITH THE BALLOON CATHETER. THE CATHETER WAS NEVER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22438042 ↗MalfunctionApr 3, 2023Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON ON A 5F 5MM X 200MM X 150 SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER POPPED IN THE VESSEL UPON INFLATION OF 8ATM OF PRESSURE. THE PROCEDURE WAS COMPLETED, AND PATIENT WAS STABLE. THERE WERE NO REPORTS OF PATIENT INJURY. THIS ISSUE OCCURRED AFTER A NON-CORDIS ATHERECTOMY DEVICE WAS USED. THE SUPERFICIAL FEMORAL ARTERY (SFA) WAS CALCIFIED. THERE WAS NO DIFFICULTY REMOVING THE PROTECTIVE BALLOON COVER, NOR WAS THERE DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. THE TARGET SITE VESSEL WAS CALCIFIED, HOWEVER THERE WAS NO TORTUOSITY AND HAD 60% STENOSIS. THERE WAS NO ACUTE ANGULATION OR BIFURCATION OF THE VESSEL. THERE WAS SOME CALCIFICATION AT THE ACCESS SITE BUT NO TORTUOSITY. THERE WAS NO STENOSIS, ACUTE ANGULATION, OR BIFURCATION OF THE ACCESS SITE VESSEL. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND AT ANY POINT DURING THE PROCEDURE. THE DEVICE WAS ABLE TO CROSS THE LESION. THE DEVICE DID NOT KINK AT ANY TIME DURING THE PROCEDURE. THERE WERE NO ANOMALIES NOTED AFTER THE DEVICE WAS DELIVERED TO THE LESION. THE DEVICE WAS ABLE TO BE REMOVED EASILY FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16665237 ↗MalfunctionFeb 13, 2023Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON OF A 5MM X 25CM 150CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED DURING AN INFLATION TO 4 ATMOSPHERES (ATM) AND WAS UNABLE TO BE FILLED. AS A RESULT, A 5MM X 22CM NON-CORDIS PTA BALLOON CATHETER WAS USED AS A REPLACEMENT. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT AN 80% CHRONIC TOTALLY OCCLUDED (CTO) LESION IN THE LEFT LOWER SUPERFICIAL FEMORAL ARTERY (SFA) WHICH HAD CALCIFICATION AND VESSEL TORTUOSITY. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WAS NO DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THE BALLOON WAS PREPPED WITH A 1:1 CONTRAST TO SALINE RATIO AND THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE DEVICE WAS ABLE TO BE REMOVED EASILY FORM THE PATIENT AND REMAINED IN ONE PIECE DURING ITS REMOVAL. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16364813 ↗MalfunctionApr 25, 2022Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, A 5MM X 25CM 150CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WAS UNABLE TO CROSS THE DISTAL END OF A LESION AND, AFTER A NORMAL INFLATION OF THE BALLOON TO NOMINAL PRESSURE, THE BALLOON WAS EXTREMELY DIFFICULT TO DEFLATE. THE BALLOON WAS EVENTUALLY DEFLATED ON ITS OWN AND WAS REMOVED EASILY FROM THE PATIENT IN ONE PIECE. ANOTHER 5MM X 25CM 150CM SABER PTA BALLOON CATHETER FROM A DIFFERENT LOT WAS USED IN PLACE OF THE INITIAL SABER PTA BALLOON CATHETER; HOWEVER, DURING AN INFLATION TO NOMINAL PRESSURE, THE BALLOON OF THE SECOND 5MM X 25CM 150CM SABER PTA BALLOON CATHETER RUPTURED. THE RUPTURED 5MM X 25CM 150CM SABER PTA BALLOON CATHETER WAS THEN REMOVED AND A 5MM X 20CM SABER PTA BALLOON CATHETER WAS USED WITHOUT ISSUE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A CALCIFIED LESION IN WHICH A NON-CORDIS .014 GUIDEWIRE WAS USED AND THERE WAS NO OBSERVED DAMAGE TO THIS GUIDEWIRE. BOTH OF THE 5MM X 25CM 150CM SABER PTA BALLOON CATHETERS WERE STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THERE WAS NO DIFFICULTY REMOVING THE DEVICES FROM THEIR HOOP, REM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14192187 ↗MalfunctionApr 25, 2022Deflation Problem; Deflation Problem
AS REPORTED, A 5MM X 25CM 150CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WAS UNABLE TO CROSS THE DISTAL END OF A LESION AND, AFTER A NORMAL INFLATION OF THE BALLOON TO NOMINAL PRESSURE, THE BALLOON WAS EXTREMELY DIFFICULT TO DEFLATE. THE BALLOON WAS EVENTUALLY DEFLATED ON ITS OWN AND WAS REMOVED EASILY FROM THE PATIENT IN ONE PIECE. ANOTHER 5MM X 25CM 150CM SABER PTA BALLOON CATHETER FROM A DIFFERENT LOT WAS USED IN PLACE OF THE INITIAL SABER PTA BALLOON CATHETER; HOWEVER, DURING AN INFLATION TO NOMINAL PRESSURE, THE BALLOON OF THE SECOND 5MM X 25CM 150CM SABER PTA BALLOON CATHETER RUPTURED. THE RUPTURED 5MM X 25CM 150CM SABER PTA BALLOON CATHETER WAS THEN REMOVED AND A 5MM X 20CM SABER PTA BALLOON CATHETER WAS USED WITHOUT ISSUE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A CALCIFIED LESION IN WHICH A NON-CORDIS .014 GUIDEWIRE WAS USED AND THERE WAS NO OBSERVED DAMAGE TO THIS GUIDEWIRE. BOTH OF THE 5MM X 25CM 150CM SABER PTA BALLOON CATHETERS WERE STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THERE WAS NO DIFFICULTY REMOVING THE DEVICES FROM THEIR HOOP, REM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14192178 ↗MalfunctionApr 23, 2022Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON OF A SABER 5MM25CM 150 RUPTURED BELOW BURST PRESSURE WHILE IN THE VESSEL. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE BALLOON WAS NOTED TO RUPTURE BELOW RBP WHILE IN THE VESSEL. THE BALLOON DID NOT RUPTURE WHILE IN A STENT. THE PRODUCT WAS STORED PROPERLY ACCORDING TO THE INSTRUCTION FOR USE (IFU). THE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP OR REMOVING THE PROTECTIVE BALLOON COVER. THERE WAS NO DIFFICULTY REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR DAMAGES NOTED PRIOR TO INSERTING INTO THE PATIENT. THE DEVICE WAS PREPPED PER THE IFU AND WAS ABLE TO MAINTAIN NEGATIVE PRESSURE. THE INTENDED PROCEDURE WAS TO TREAT THE SUPERFICIAL FEMORAL ARTERY. THE LESION WAS NOTED TO BE CALCIFIED. THE VESSEL DID NOT HAVE ANY TORTUOSITY. THE DEVICE WAS NOT BEING USED TO TREAT A CHRONIC TOTAL OCCLUSION. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE DEVICE THROUGH THE ROTATING HEMOSTATIC VALVE. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON THROUGH THE GUIDE CATHETER. THE CATHETER ADVANCED THROUGH THE VESSEL WITHOUT DIFFICULTY. THE DEVICE CROSSED THE LESION WITHOUT DIFFICULTY. THE CATHETER WAS NE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14191020 ↗MalfunctionMar 23, 2022Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON OF A 5MMX 25CM 150CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED DURING AN INFLATION TO 14 ATMOSPHERES (ATM). AS A RESULT, THE SABER PTA BALLOON CATHETER WAS REMOVED AND A 5MM X 20CM NON-CORDIS BALLOON CATHETER WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT ARTERIOSCLEROSIS OBLITERANS IN THE SUPERFICIAL FEMORAL ARTERY (SFA). THE LESION IN THE SFA WAS REPORTED AS A LONG SEGMENT STENOSIS OF 90% WHICH WAS CONFIRMED TO BE A FUNCTIONAL CHRONIC TOTAL OCCLUSION (CTO). THERE WAS CALCIUM PRESENT AT THE TARGET LESION; HOWEVER, THERE WERE NO SIGNS OF TORTUOSITY TO THE SFA. THE DEVICE WAS PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WERE NO DIFFICULTIES REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION AND THERE WERE NO KINKS OR OBSERVED DAMAGES NOTED PRIOR TO INSERTION. A NON-CORDIS .018 GUIDEWIRE WAS USED FOR THE PROCEDURE. THERE WERE NO DAMAGES OBSERVED ON THE .018 GUIDEWIRE AND THE WIRE WAS SUCCESSFULLY DELIVERED TO THE LESION FOLLOWED BY THE INSERTION OF THE SABER PTA BALLOON CATHETER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13863450 ↗MalfunctionSep 9, 2020Material Rupture; Material Rupture
DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA, THE PROXIMAL TIP OF A 5MM X 25CM X 150CM SABER PTA BALLOON CATHETER LEAKED DURING INFLATION AND THE BALLOON BURST AFTER A MAXIMUM INFLATION WAS 8 ATMOSPHERES (ATM). THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY USING ANOTHER BALLOON. THE 60% OCCLUDED LESION IN THE WAS MODERATELY CALCIFIED WITH NO VESSEL TORTUOSITY. THE ACCESS SITE WAS THE FEMORAL ARTERY. THE PROCEDURE BEING PERFORMED WAS A FEMORAL ARTERY STENOSIS OPENING. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE FORMING TUBE, NO DIFFICULTY REMOVING THE PROTECTIVE SHEATH FROM THE NEEDLE AND NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT THE PRODUCT INTO THE PATIENT. THE DEVICE PREPPED NORMALLY. THE CONTRAST USED WAS BY GE WITH A MERIT INFLATION DEVICE, AT A 1:1 CONTRAST TO SALINE RATIO. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE AND NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON THROUGH THE GUIDE CATHETER. THERE WAS NO DIFFICULTY ADVANCING THE BALLOON CATHETER THROUGH THE VESSEL AND NO DIFFICULTY CROSSING THE LESION. THE
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10507290 ↗
FDA clearance and classification
Cleared through 510(k) K133843 on Jun 27, 2014 (decision: Substantially Equivalent). Submitted by Cordis Corporation as “Saber Percutaneous Transluminal Angioplasty Dilatation Catheter”. The FDA review took 191 days. FDA record ↗
The FDA regulates product code LIT (catheter, angioplasty, peripheral, transluminal) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K133843. One submission often covers many models, so check the product and lot details in each record.
- Z-2235-2021 · Jul 19, 2021 · TerminatedCordis SABER PTA Balloon Dilation Catheter, 4F, REF 48004010X
Common questions about the SABER 48005025X
Has the SABER 48005025X been recalled?
No FDA recall record in our data names this exact device identifier (DI 20705032069561), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the SABER 48005025X?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Sep 9, 2020 and Jul 8, 2025. In those reports, the most common specific device problems are “Burst Container or Vessel” (6), “Deflation Problem” (1) and “Material Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SABER 48005025X?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SABER 48005025X FDA approved?
It was cleared through 510(k) K133843 on Jun 27, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the SABER 48005025X, and is it still sold?
The FDA's device database lists Cordis Corporation as the company responsible for it (brand SABER, model 48005025X). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal880 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal439 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal383 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal280 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal128 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page