Skip to content

SABER 48002006XRecalls & reported problems

Cordis US Corp. · SABER 2MM6CM 150 · Class 2

All models of SABER (51)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032069912), received between Jun 13, 2024 and Jan 3, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K133843); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, A 2MM 6CM SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOONS THAT IS FAULTY. THE DEVICE WAS OPENED, AND THE TIP HAD COME OFF SO IT WAS NOT USED INSIDE THE PATIENT / DURING THE PROCEDURE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21067167 ↗
  • MalfunctionJun 13, 2024Deflation Problem; Deflation Problem

    AS REPORTED, THE OPERATOR CHOSE A BALLOON 2*60MM SABER BALLOON FOR DILATATION. AFTER DILATATION, THE BALLOON WAS NOT EASILY DEFLATED TO THE ORIGINAL STATE. THE OPERATOR DID THE SECOND INFLATION, AND THE BALLOON COULD NOT BE INFLATED COMPLETELY. THE PRODUCT WAS THEN WITHDRAWN, AND THE PROCEDURE WAS COMPLETED USING OTHER DEVICES. THERE WAS NO REPORTED INJURY TO THE PATIENT. A RETROGRADE PUNCTURE WAS MADE TO THE RIGHT FEMORAL ARTERY. AN ANTERIOR FEMORAL ARTERY OCCLUSION WAS SEEN ON IMAGING. THE DEVICE WILL BE RETURNED FOR EVALUATION. ADDITIONAL INFORMATION WAS REQUESTED; HOWEVER, THE INFORMATION WAS NOT OBTAINED AFTER MULTIPLE ATTEMPTS. ADDENDUM: PRODUCT EVALUATION REVEALED A DETACHED PIECE OF THE INNER BODY OBSTRUCTING THE GUIDEWIRE LUMEN OF THE SABER BALLOON CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19533183 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133843. One submission often covers many models, so check the product and lot details in each record.

  • Z-2235-2021 · Jul 19, 2021 · Terminated
    Cordis SABER PTA Balloon Dilation Catheter, 4F, REF 48004010X

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page