Skip to content

Saber .035 48007004RRecalls & reported problems

Cordis Corporation · SABER .035 PTA Dilation Catheter 7mm x 40mm x 135cm · Class 2

All models of Saber .035 (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032080122), received between Sep 15, 2025 and Sep 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, A 7X40 .035 SABER PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON RUPTURED AT NOMINAL PRESSURE AFTER BEING INFLATED INSIDE OF A STENT. THE BALLOON RUPTURED BELOW THE RATED BURST PRESSURE (RBP). THE PROCEDURE WAS COMPLETED WITH POST-DILATATION USING A SABER .035 PTA BALLOON. NO HARM OR INJURY OCCURRED TO THE PATIENT. THE PRODUCT WAS STORED PROPERLY PER THE INSTRUCTIONS FOR USE (IFU). THE BALLOON WAS REMOVED FROM THE HOOP, PROTECTIVE COVER, AND STERILE PACKAGING WITHOUT DIFFICULTY. THERE WERE NO KINKS OR DAMAGES WERE NOTED PRIOR TO INSERTION. THE DEVICE WAS PREPPED ACCORDING TO THE IFU AND MAINTAINED NEGATIVE PRESSURE NORMALLY. THE INTENDED PROCEDURE TARGETED A LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA), WHICH WAS SEVERELY CALCIFIED WITH NO VESSEL TORTUOSITY AND 70¿80% STENOSIS. THE DEVICE WAS NOT USED FOR A CHRONIC TOTAL OCCLUSION (CTO). NO RESISTANCE OR FRICTION OCCURRED DURING INSERTION THROUGH THE ROTATING HEMOSTATIC VALVE OR WHILE ADVANCING THROUGH THE VESSEL OR CROSSING THE LESION. THE CATHETER WAS NOT IN AN ACUTE BEND AND DID NOT KINK DURING USE. A 50/50 CONTRAST-TO-SALINE RATIO WAS USED. THE AFFECTED UNIT WAS DISCARDED BY THE STAFF POST PROCEDURE AND IS N

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23054843 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page