Skip to content

Saber .035 48007010RRecalls & reported problems

Cordis Corporation · SABER .035 PTA Dilation Catheter 7mm x 100mm x 135cm · Class 2

All models of Saber .035 (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032080146), received between Feb 23, 2023 and May 13, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2025Leak/Splash; Leak/Splash

    AS REPORTED, WHEN FLUSHING A 7MM X 100MM SABER .035 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER WITH A SYRINGE, SALINE SQUIRTED OUT OF THE PROXIMAL PORTION OF THE BALLOON. AN UNKNOWN BALLOON CATHETER WAS USED AS A REPLACEMENT TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. THE DEVICE WILL BE RETURNED FOR EVALUATION. ADDENDUM: PRODUCT EVALUATION REVEALED A LEAKAGE THROUGH THE BODY/SHAFT OF THE SABER PTA BALLOON CATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22013425 ↗
  • InjuryFeb 23, 2023Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component; Burst Container or Vessel; Difficult to Remove; Detachment of Device or Device Component

    AS REPORTED, THE BALLOON OF A 7MM X 10CM 135CM SABER .035 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER POPPED DUE TO SEVERE CALCIFICATION OF THE MID TO DISTAL SUPERFICIAL FEMORAL ARTERY (SFA). NEGATIVE PRESSURE WAS THEN APPLIED TO THE DEVICE WITH AN INDEFLATOR SYRINGE AND BLOOD FILLED THE SYRINGE CONFIRMING THE RUPTURE. AFTER APPLYING NEGATIVE PRESSURE, THE DEVICE WAS RETRACTED THROUGH AN 8F CORDIS CATHETER SHEATH INTRODUCER (CSI). WHEN THE BALLOON WAS REMOVED, IT WAS NOTICED THAT THE ACTUAL BALLOON COATING WAS MISSING FROM THE BALLOON CATHETER. THE PHYSICIAN WAS UNABLE TO SEE THE SHEARED PORTION OF BALLOON UNDER FLUOROSCOPY AND THERE WAS CONCERN THAT THE BALLOON WAS SHEARED OFF THE BALLOON CATHETER WHEN RETRACTED INTO THE 8F UNKNOWN CORDIS CSI. THEREFORE, IT WAS DECIDED TO INTENTIONALLY CUT THE 8F CSI INTO MULTIPLE PIECES TO SEE IF THE SEPARATED BALLOON SEGMENT WAS WITHIN THE 8F CSI. THE SEPARATED BALLOON PORTION WAS EVENTUALLY OBSERVED INSIDE OF THE PATIENT. MULTIPLE ATTEMPTS WERE THEN MADE TO SNARE THE REMAINING BALLOON SEGMENT AND REMOVE IT FROM THE PATIENT BUT WERE UNSUCCESSFUL. AS A RESULT, A 7MM X 150MM SMART SELF-EXPANDING STENT (SES) AND AN 8MM X 40MM SMART

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16432625 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →