Skip to content

Saber .035 48012004RRecalls & reported problems

Cordis Corporation · SABER .035 PTA Dilation Catheter 12mm x 40mm x 135cm · Class 2

All models of Saber .035 (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032080634), received between Jul 17, 2024 and Jul 17, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Frayed” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Frayed1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 17, 2024Material Frayed; Material Split, Cut or Torn; Material Frayed; Material Split, Cut or Torn

    AS REPORTED, THE BALLOON OF A 12MM X 40MM SABER .035 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER DID NOT FILL WHEN INSIDE OF THE PATIENT. THE OPERATOR WITHDREW THE DEVICE AND ATTEMPTED TO INFLATE THE BALLOON OUTSIDE OF THE PATIENT WHERE A SMALL HOLE IN THE BODY OF THE BALLOON WAS OBSERVED BUBBLING FLUID. THE PROCEDURE WAS ENDED AND THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO DILATE A DIALYSIS ACCESS STENOSIS IN A SUPERIOR VENA CAVA (SVC) WHICH HAD A 70% SEPTAL STENOSIS. AN APPROACH FROM THE RIGHT INTERNAL JUGULAR VEIN WAS MADE FOR THIS PROCEDURE. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WAS NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. THE DEVICE WAS PREPPED WITH A 50/50 CONTRAST AND SALINE MIXTURE AND MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE DEVICE WAS ABLE TO BE REMOVED EASILY FROM THE PATIENT AND REMAINED IN ONE PIECE DURING ITS REMOVAL. THE OPERATOR IS TRAINED AND EXPERIENCED IN THE USE OF BALLOON CATHETERS. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19765566 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates