Skip to content

SaberX Radianz R51005003LRecalls & reported problems

Cordis US Corp. · SABERX RADIANZ 5MM3CM 190 · Class 2

All models of SaberX Radianz (24)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032081839), received between Oct 17, 2023 and Oct 17, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 17, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, THE DR. INTENDED TO USE A 5MM X 3CM X 190CM SABERX RADIANZ TO TREAT AN IN-STENT RESTENOSIS OF A STENT IN THE LEFT PROXIMAL EXTERNAL ILIAC ARTERY. HE USED A STANDARD 5FR UNKNOWN SHEATH TO GAIN ACCESS IN THE BRACHIAL ARTERY. IT WAS SUGGESTED TO USE A CORDIS BRITE TIP RADIANZ TO SUPPORT THE SABERX RADIANZ BALLOON, BUT HE DID NOT FEEL IT WOULD BE NECESSARY. THE DR WAS ABLE TO GET THE BALLOON DOWN CLOSE TO THE LESION SIGHT ALTHOUGH IT DID NOT APPEAR VERY SMOOTHLY. AT ONE POINT THE NON-CORDIS .018 GUIDEWIRE IS THOUGHT TO HAVE PROLAPSED UP INTO THE PATIENT'S AORTA. IT WAS OBSERVED THEY WERE HAVING SOME UNUSUAL ARRHYTHMIAS. THE ARRYTHMIA DID CORRECT ITSELF. ONCE THE BALLOON WAS CLOSE TO THE LESION SIGHT THE DR. THEN BEGAN TO HAVE DIFFICULTY PULLING IT BACK INTO THE CORRECT POSITION DIRECTLY OVER THE LESION. DUE TO THIS DIFFICULTY, THE DR. DECIDED TO REMOVE THE BALLOON TO SEE IF THERE WAS SOMETHING WRONG WITH IT. AS HE WAS PULLING THE BALLOON BACK TO THE SHEATH, HE BEGAN HAVING PROGRESSIVELY MORE DIFFICULTY PULLING IT BACK. HE CONTINUED PULLING THE BALLOON BACK AND THEN IT WAS OBSERVED TO HAVE BROKEN OFF AT THE BALLOON CATHETER SHAFT, WITH PART OF THE BALLOON SHAFT STILL STUCK

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17951485 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page