SABER .014 PTA Dilatation Catheter 47020150MRecalls & reported problems
Cordis US Corp. · PTA, OTW, 2.0 x 150, 150cm · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032087435), received between Mar 16, 2026 and Mar 16, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 16, 2026Burst Container or Vessel
AS REPORTED, THE BALLOON OF A 2MM X 150MM SABER .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED AT NOMINAL PRESSURE. ANOTHER UNKNOWN SABER .014 PTA BALLOON CATHETER WAS USED AS A REPLACEMENT. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU), AND THERE WAS NO DIFFICULTY REMOVING THE STERILE PACKAGING COMPONENTS OR THE PROTECTIVE BALLOON COVER. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION, AND A 50/50 CONTRAST-TO SALINE RATIO WAS USED. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND AND DID NOT KINK AT ANY TIME DURING THE PROCEDURE. IT WAS REMOVED EASILY FROM THE PATIENT AND WAS REMOVED INTACT IN ONE PIECE. ADDITIONAL DETAILS WERE REQUESTED BUT WERE UNKNOWN. THE DEVICE WAS DISCARDED AND WILL NOT BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24612175 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page