Skip to content

SABER .014 PTA Dilatation Catheter 47015040MRecalls & reported problems

Cordis US Corp. · PTA, OTW, 1.5 x 40, 150cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032087688), received between Aug 27, 2024 and Aug 27, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2024Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, THE 1.5 X 40 SABER .014 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON WAS OPENED AND INSERTED INTO THE PATIENT. WHEN THE BALLOON WENT UP IN THE ARTERY, THE PRODUCT BURST BELOW RBP ON PRESENT CALCIUM.THE DEVICE WAS REMOVED INTACT FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT UNKNOWN BALLOON AND THE CTO WAS EVENTUALLY CROSSED AND COMPLETED. THE PATIENT WAS UNHARMED. THE DEVICE WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, STYLET, OR ANY OTHER STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE DEVICE INTO THE PATIENT. THE DEVICE WAS PREPPED PER THE IFU AND WAS ABLE TO MAINTAIN NEGATIVE PRESSURE. THE INTENDED LESION WAS NOTED TO HAVE A CHRONIC TOTAL OCCLUSION (CTO). THERE WAS NO RESISTANCE FORWARD SLASH FRICTION WHILE INSERTING THE DEVICE THROUGH ROTATING HEMOSTATIC VALVE OR WHILE INSERTING THROUGH THE UNKNOWN GOD CATHETER. THERE WAS DIFFICULTY NOTED WHILE ADVANCING THROUGH THE VESSEL DUE TO THE CTO. THE BALLOON CATHETER DID NOT KINK WHILE BEING USED. THE DEVICE WILL NOT BE RETURNED FOR EVA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20079464 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page