SABER .014 PTA Dilatation Catheter 47030300MRecalls & reported problems

Cordis US Corp. · PTA, OTW, 3.0 x 300, 150cm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032089095), received between Nov 17, 2023 and May 6, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Leak/Splash1
  • Material Frayed1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, THE BALLOON OF A SABER .014 PTA OTW, 3.0 X 300, 150CM PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER SEPARATED FROM ITS SHAFT AND RAIL DURING REMOVAL FROM THE PATIENT. A NON-CORDIS SNARE WAS USED TO REMOVE THE SECTION OF BALLOON THAT REMAINED OVER THE UNKNOWN WIRE IN THE PATIENT. THE BALLOON WAS REMOVED SUCCESSFULLY, AND NO HARM WAS CAUSED TO THE PATIENT. A CONTRALATERAL APPROACH WAS USED. THE TARGET SITE VESSEL WAS CALCIFIED, NO TORTUOSITY, NO ACUTE OR BIFURCATING ANGLE. THE VESSEL CHARACTERISTICS ACCESSING TARGET SITE HAD NO CALCIFICATION, TORTUOSITY, NO ACUTE OR BIFURCATING ANGLE. THERE WAS NO DIFFICULTY REMOVING THE PROTECTIVE BALLOON COVER. THERE WAS NO DIFFICULTY REMOVING ANY OF THE STERILE PACKAGING COMPONENTS. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THE DEVICE WAS NOT PUT INTO AN ACUTE BEND AT ANY POINT DURING THE PROCEDURE. THE DEVICE WAS ABLE TO CROSS THE LESION, HOWEVER WAS CAUGHT IN IT. THE BALLOON COULD HAVE POTENTIALLY BEEN CAUGHT IN A DEPLOYED STENT, AS THE SECTION OF THE BALLOON THAT BECAME SEPARATED WAS FOUND DISTAL TO AN EXISTING STENT. THE DEVICE DID NOT KINK AT ANY TIME DURING THE PROCEDURE. THERE WERE NO ANOMALIES

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21971889 ↗
  • MalfunctionNov 17, 2023Material Frayed; Leak/Splash; Material Split, Cut or Torn; Material Frayed; Leak/Splash; Material Split, Cut or Torn

    AS REPORTED, THE 3X300 SABER .014 BALLOON EXTENDS SIGNIFICANTLY PAST THE PROXIMAL MAKERS AND SEEM TO MOVE, OR ¿DANCE¿, WITH INCREASING PRESSURE APPLIED. THE DEVICE WAS REMOVED FROM THE PATIENT INTACT. THE CASE WAS COMPLETED WITH A NEW 3X300 SABER .014 BALLOON. THERE WAS NO REPORTED INJURY TO THE PATIENT. THE PRODUCT WAS STORED PROPERLY ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, STYLET, OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE DEVICE WAS PREPPED NORMALLY AND WAS ABLE TO MAINTAIN NEGATIVE PRESSURE. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON THROUGH THE GUIDE CATHETER. THE CATHETER WAS NEVER IN AN ACUTE BEND AND WAS NEVER KINKED DURING USE. THERE WAS NO DIFFICULTY ADVANCING THE BALLOON CATHETER THROUGH THE VESSEL. THERE WAS NO DIFFICULTY CROSSING THE LESION. THE LESION WAS IN THE TIBIAL ARTERY. THE LESION WAS NOTED TO HAVE MINIMAL CALCIFICATION. THE VESSEL DID NOT HAVE ANY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18160851 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →