Harmonic HPBLUERecalls & reported problems

Ethicon Endo-Surgery, LLC · Harmonic Blue Hand Piece and Blue Test Point · Class U

Quick summary

This Ethicon Harmonic Blue ultrasonic surgical handpiece has 133 reports, many describing the device not responding, showing error messages or overheating. The reports mention 7 injuries, including some burns, and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
133
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 133 adverse event reports that list this exact device (DI 20705036002151), received between Dec 12, 2017 and Jun 11, 2026 — classified as 7 injuries, 126 malfunctions.

The most frequently reported specific device problem is “Failure to Read Input Signal” (48 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 6 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 4 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Read Input Signal48
  • Insufficient Device Problem Information21
  • Difficult or Delayed Activation13
  • Excessive Heating10
  • Protective Measures Problem7
  • Connection Problem5
  • Adverse Event Without Identified Device or Use Problem4
  • Break4

Patient problems

  • No Clinical Signs, Symptoms or Conditions96
  • No Consequences Or Impact To Patient31
  • Burn(s)4
  • Hemorrhage/Blood Loss/Bleeding2
  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 25 most recent of 133 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2026Difficult or Delayed Activation

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY THE BLUE HANDPIECE DID NOT WORKED. WHEN THE HANDPIECE WAS CHANGED THE DEVICE WORKED. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25469701 ↗
  • MalfunctionJun 11, 2026Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THAT THE HPBLUE GAVE ERRORS. BY TESTING THE DEVICE IT SEEMS TO BE DEFECT WHILE IT STILL HAS 29 REMAINING USES. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25469502 ↗
  • MalfunctionMar 16, 2026Failure to Read Input Signal

    IT WAS REPORTED THAT DURING A CABG, WHILE ASSEMBLING WITH SINGLE USE DEVICE, IT "WOULD NOT WORK." NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24605788 ↗
  • MalfunctionMar 16, 2026Activation Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE BLUE HAND PIECE IS BROKEN. IT REQUIRES TO MAKE THE TEST AGAIN AND AGAIN. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24605726 ↗
  • MalfunctionDec 24, 2025Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THERE WAS A REPLACE INSTRUMENT ERROR MESSAGE. THERE WAS NO PATIENT CONSEQUENCES IDENTIFIED WHEN THIS COMPLAINT EVENT WAS CAPTURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23887911 ↗
  • MalfunctionNov 12, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    THE DEVICE WAS RECEIVED WITH NO EVENT DETAILS. ATTEMPTS WERE MADE TO OBTAIN DETAILS AND NO INFORMATION WAS PROVIDED. NO PATIENT CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23536290 ↗
  • MalfunctionAug 21, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WAS NOT WORKING. NO OTHER INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22861793 ↗
  • MalfunctionMay 6, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE HANDPIECE OVERHEATS ABNORMALLY WHEN USED. IT WAS SEEN WITH THE UROLOGY OPERATING ROOM. THE FORCEPS HAVE 33 USES REMAINING. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21971342 ↗
  • MalfunctionApr 23, 2025Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THERE ARE STILL 25 TIMES LEFT TO USE BUT HAND PIECE DIDN'T WORKED. THEY TRIED TO USE IT SEVERAL TIMES BUT NOT WORKING. THE PROCEDURE WAS DELAYED 10 MINUTES BUT WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21892167 ↗
  • MalfunctionMar 26, 2025Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT, DURING AN UNKNOWN PROCEDURE, THE DEVICE WAS NOT RECOGNIZED. NO PIECES WERE FELL INTO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21698119 ↗
  • MalfunctionMar 3, 2025Failure to Read Input Signal; Difficult or Delayed Activation; Failure to Read Input Signal; Difficult or Delayed Activation

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THERE WAS A HPBLUE FAILURE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21506349 ↗
  • MalfunctionFeb 5, 2025Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE AN ERROR MESSAGED SHOWED: "REACTIVATE. TWO SWITCH ACTIVATIONS DETECTED: MANUAL CONTROL, REACTIVATE INSTRUMENT TO CONTINUE." THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21314562 ↗
  • MalfunctionFeb 5, 2025Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE OR NURSE TRIED CONNECTING A HARMONIC TO THE BLUE HANDPIECE AND AN ERROR MESSAGE SHOWED: REPLACE HANDPIECE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21314546 ↗
  • MalfunctionFeb 4, 2025Noise, Audible; Noise, Audible

    IT WAS REPORTED THAT DURING A TRAUMA PROCEDURE, WHEN USING THE DEVICE, THERE HAD BEEN A VERY LOUD VOICE IN HAND PIECE. THEY CAN NOT EXPLAIN IT BETTER BUT VERY UNNORMAL SOUND. STILL 46 USING TIMES LEFT. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21306057 ↗
  • MalfunctionFeb 4, 2025Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THERE WAS A DEFECTIVE HAND-PIECE. 40 USES REMAINING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21304947 ↗
  • MalfunctionDec 10, 2024Break; Break

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE IT WAS FOUND THAT THE HANDPIECE SCREW IS DAMAGED AT THE TIP BEFORE USE ON PATIENT DURING PHYSICAL INSPECTION. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20896232 ↗
  • MalfunctionDec 9, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING A THYROIDECTOMY THERE WAS A DEFECTIVE DEVICE - THE CABLE DOESN'T WORK ANYMORE. NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20885616 ↗
  • MalfunctionOct 21, 2024Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE HAND PIECE ANNOUNCED ERROR DURING GETTING TOGETHER WITH INSTRUMENT BUT WITH USING TESTING STICK NO PROBLEM FIRST AND SECOND TIME BUT THIRD TIME ANNOUNCE ERROR - REPLACE HAND PIECE. COSTUMER USE SEVERAL NEW INSTRUMENTS. HAND PIECE HAVE 30 USAGES YET. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20493085 ↗
  • MalfunctionSep 20, 2024Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WASN'T ABLE TO SEAL AND CUT. THERE WERE NO VISIBLE ISSUES. THE HANDPIECE HAD 7 USES LEFT. THE DISPOSABLE AND HANDPIECE WERE REPLACED AND IT WORKED FINE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20266096 ↗
  • MalfunctionSep 4, 2024Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE HANDPIECE DID NOT WORK ANYMORE, WHILE THERE ARE MORE THAN 90 USES LEFT ON ONE. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20136023 ↗
  • MalfunctionAug 5, 2024Crack; Failure to Read Input Signal; Connection Problem; Crack; Failure to Read Input Signal; Connection Problem

    IT WAS REPORTED THAT DURING A THYROIDECTOMY THE SURGEON WAS USING THIS HPBLUE WITH HAR9F AT FIRST WHICH IT WENT WELL. SUDDENLY THE HAR9F COULD NOT BE ACTIVATED. THUS HE DECIDED TO CHANGE ANOTHER CABLE WITH THE SAME HAR9F TO CONTINUE THE SURGERY. THE PROCEDURE WAS DELAYED. THE NURSE TRIED TO USE THIS CABLE WITH THE SAME HAR9F AFTER CASE, WHICH GIVING SAME ERROR MESSAGE. SALESMAN WENT ON-SITE TO CHECK THE CABLE AND FOUND THERE'S ABNORMAL SITUATION WITH THE CABLE. THE 7 PINS WHERE TO CONNECT WITH ADAPTER GO DARKEN (LOOK LIKE COVERING WITH BLACK DUST) WHICH IS DIFFERENT FROM THE SHINNY PINS WE USUALLY SEE WITH CABLE. SALESMAN ALSO FOUND THERE'S A BLACK METALLIC/PLASTIC FRAGMENT AT THE CONNECTING AREA WITH CONSUMABLES. SALESMAN TRIED TO CONNECT IT WITH GEN11 AND UNDERWENT HANDPIECE TESTING, HOWEVER IT POPPED OUT WITH 'REACTIVE' 'TWO SWITCH ACTIVATING DETECTED' AND COULD NOT PASS THE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19911495 ↗
  • MalfunctionJun 25, 2024Crack; Crack

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE CUSTOMERS HPBLUE/HAR9F IS NOT PASSING THE ACTIVATION TEST. REP SAID THE CUSTOMER TRIED MULTIPLE GENERATORS WITH SAME HANDPIECE AND 2 HAR9F DEVICES AND COULD NOT GET EITHER TO ADVANCE PAST THE ACTIVATION SCREEN. CHECKED THE HANDPIECE THEY REPORTED AN ISSUE WITH AND THE CONTACT RING IS FILTHY WHICH MAY BE THE PROBLEM. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19609857 ↗
  • MalfunctionJun 21, 2024Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN ENT PROCEDURE THAT THE FIRST HARMONIC WAS NOT SEALING PROPERLY. A SECOND HARMONIC WAS OPENED AND THE GENERATOR GAVE THE ERROR TO REPOSITION JAWS. THE HANDPIECE AND GENERATOR WAS REPLACED BUT THE ISSUE STILL PERSISTED. A COMPETITOR'S DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19582826 ↗
  • MalfunctionJun 7, 2024Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING A THYROIDECTOMY AFTER CAREFUL SETUP IT WORKS THEN ERROR MESSAGES KEEP COMING UP - TO TIGHTEN THE HAND PIECE - WHICH THE STAFF DO AND THEN IT WORKS FOR A WHILE AND THEN FURTHER ERROR MESSAGES COME UP. 80 USES REMAINING. ANOTHER HANDPIECE OPENED THE PROCEDURE WAS DELAYED BUT COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19491344 ↗
  • MalfunctionJun 7, 2024Failure to Read Input Signal; Failure to Read Input Signal

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WAS IN ERROR MODE.NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19490727 ↗

Other Instrument, ultrasonic surgical devices

All Instrument, ultrasonic surgical devices →